11.0 ERROR CODES


During the initial power-on testing or during operation, the scale may show an
error message. The meaning of the error messages is described below.

If an error message is shown, repeat the step that caused the message,
turning the balance on, carry out the calibration or other functions. If the error
message is still shown contact your dealer for further support.

ERROR CODE

DESCRIPTION

POSSIBLE CAUSES

Err 1

Time input error.

Tried to set an illegal time, i.e. 26hours

Err 2

Date input error

Tried to set an illegal date, i.e. 36

th

day

Err 4

Initial Zero is greater than
allowed (typically 4% of
the maximum capacity)
when power is turned on
or when the

[Zero] key is

pressed,

Weight is on the pan when turning the
scale on.

Excessive weight on the pan when
zeroing the scale.

Improper calibration of the scale.

Damaged load cell.

Damaged Electronics.

Err 6

A/D count is not correct
when turning the scale
on.

Platform is not installed.

Damaged Load cell.

Damaged Electronics.