8.

Mix the reactions gently by pipetting up and down several times with a pipettor.

9.

Cover the plate and incubate for 30 minutes at room temperature on a plate shaker.

10. Clip the plate in place on the Luminex Magnetic Plate Separator and rapidly and

forcefully invert over a biohazard receptacle to evacuate the liquid from the wells.

NOTE: For information on the MagPlex Manual Wash Method, please visit:

http://www.luminexcorp.com/Products/ReagentsMicrospheres/MAGPLEX-
MICROSPHERES.

11. Wash each well with 100 μL of PBS-1% BSA by gently pipetting up and down

several times with a pipettor, and remove the liquid by using the procedure
described in the previous step.

12. Repeat the previous wash step once.
13. Resuspend the microspheres in 100 μL of PBS-1% BSA by gently pipetting up and

down several times with a pipettor.

14. Analyze 50-75 μL on the Luminex analyzer according to the system manual. An

example of typical results is shown below.