beautypg.com

Laurel Electronics LTM DPM SERIES ANALOG INPUT, SERIAL DATA OUTPUT User Manual

Page 38

background image

- 38 -

Load Cell & Microvolt Input

Range

Input Resistance

Input Error

±20.000 mV
±50.000 mV
±100.00 mV
±250.00 mV
±500.00 mV

1 GΩ

0.01% of FS

±2 counts

Max applied voltage ..............................................................................................................100 V

RTD Input

(1°, 0.1° or 0.01° resolution)

Type

Excitation

Range

Conformity Error

Platinum, Pt100

=.00385 (DIN)

256 µA

-202 to 850°C

-331 to 1562°F

0.03°C

0.05°F

Platinum, Pt100

=.003925 (ANSI)

256 µA

-202 to 631°C

-331 to 1168°F

0.04°C

0.07°F

Nickel, Ni120

=.00672

256 µA

-80°C to +260°C

-112°F to +500°F

±0.05°C

±0.09°F

Copper, Cu10

=.00427

5.0 mA

-97°C to +260°C

-143°F to +500°F

±0.05°C

±0.09°F

RTD connection: ......................................................................................................2, 3 or 4 wire
Span tempco.............................................................................................. ±0.003% of reading/ºC
Zero tempco ........................................................................................................... ±0.03 deg/deg
Sensor lead resistance tempco per conductor, 2-wire....................10 mdeg / Ω / deg up to 10 Ω
Sensor lead resistance tempco per conductor, 3 & 4-wire ...........10 mdeg / Ω / deg up to 100 Ω
Over-voltage protection .................................................................................................... 125 Vac
Open sensor indication ..............................................0 mA or > 20 mA output, jumper selectable

Thermocouple Input

(1° or 0.1° resolution)

Selection of signal ranges & temperature sensors ................................Via jumpers and software
Selection of serial format, excitation output, analog output................ Via jumpers on main board
Programming........................................ Via PC using Instrument Setup software and serial cable
Input resistance .................................................................................................................... 1 GΩ
Input current.......................................................................................................................100 pA
Overall input accuracy .................................................................. ±

0.01% of full span ±2 counts

Max lead resistance ..........................................................................1 kΩ max for rated accuracy
Span tempco.............................................................................................. ±0.003% of reading/ºC
Reference junction tempco ..................................................................................... ±0.02 deg/deg