Click on

DPM

=>

Get Setup

to retrieve the current setup information from your DPM trans-

mitter, then on

View

=>

Setup

, which will take you to the

Input+Display

tab. Use this screen to

set up

Signal Input

,

Display

, and

Control Inputs

. The software reads the signal conditioner type,

but not the range, which is set by jumpers.

50/60 Hz Line Freq

is used for noise rejection.

Click on the

Scaling

tab to scale your transmitter. You will be given the choice of three scaling

methods: 1) Scale and Offset method, 2) Coordinates of 2 points method where (Low In, Low
Read) and (High In, High Read) data points are entered numerically, and 3) Reading Coor-
dinates of 2 points method, which captures actual readings.