D100 Series ASCII Command Set

Command and Definition

Typical

Typical

Command

Response

Message Message

($ prompt)

DI

Digital Input

$1DI

*0003

DO

Digital Output

$1DOFF

*

RD

Read Data

$1RD

*+00072.00

RS

Read Setup

$1RS

*31070142

RSU

Read Setup

$1RSU

*31070142

RZ

Read Zero

$1RZ

*+00000.00

WE

Write Enable

$1WE

*

Write Protected Commands

CZ

Clear Zero

$1CZ

*

RR

Remote Reset

$1RR

*

SU

Setup Module

$1SU31070142

*

TS

Trim Span

$1TS+00600.00

*

TZ

Trim Zero

$1TZ+00000.00

*

Setup

The D100 series are initiated at the factory using the Dataforth ASCII protocol.

This allows setup and configuration, including the Modbus device address, to

be easily performed using the Dataforth setup software or a dumb terminal. Each

D100 module must be properly configured before installation into a Modbus

system.

Utility Software

Complimentary Utility Software is included with each purchase order. The

software is compatible with Windows 95, 98, NT 4.0+, 2000 operating systems

and distributed on CD-ROM. The Utility Software simplifies configuration of all

user-selectable options such as device address, baud rate and filtering

constants.

Process Control Software

Modbus RTU protocol is supported by virtually all commercial process control

software programs available today. These programs operate on IBM and

compatible personal computers in the Windows 95, 98, NT and IBM OS/2

environments.

Specifications

Typical at T

A

=+25°C and nominal power supply unless otherwise noted

Analog

• Single channel analog input.

• Maximum CMV, input to output at 60Hz: 500Vrms.

• Leakage current, input to output at 115Vrms, 60Hz: <2µA rms.

• 15 bit measurement resolution.

• 8 conversions per second.

• Autozero & autocalibration—no adjustment pots.

Digital

• 8-bit CMOS microcomputer.

• Digital scaling, linearization and calibration.

• Nonvolatile memory eliminates pots and switches.

Digital filtering

• Small and large signal with user selectable time

constants from 0 to 16 seconds.

Communications

• Communications in MODBUS-RTU via RS-485 ports.

• Selectable baud rates: 300, 600, 1200, 2400, 4800,

9600, 19200, 38400.

• NRZ asynchronous data format; 1 start bit, 8 data bits,

1 parity bit and 1 stop bit.

• Parity: odd, even, none.

• User selectable channel address.

• Up to 247 multidrop modules per host serial port.

• Communications distance up to 4,000 feet (RS-485).

• Transient suppression on RS-485 communications lines.

• All communications setups stored in EEPROM.

Power

• Requirements:

Regulated +5VDC ,

0.75W max (DIN-150, 2.0W max.).

• Protected against power supply reversals.

Environmental

• Temperature Range: Operating -25°C to +70°C.

Storage -25°C to +85°C.

• Relative Humidity: 0 to 95% noncondensing.

D11x Voltage Inputs

• Voltages: ±10mV, ±100mV, ±1V, ±5V, ±10V, ±100VDC.

• Resolution: 0.01% of FS (4 digits).

• Accuracy: ±0.02% of FS max.

• Common mode rejection: 100dB at 50/60Hz.

• Zero drift: ±1 count max (autozero).

• Span tempco: ±50ppm/°C max.

• Input burnout protection to 250VAC .

• Input impedance:

±1V input = 100M

min.

±5V input = 1M

min.

D125 Current Inputs

• Currents: 4-20mADC.

• Resolution: 0.04% of FS.

• Accuracy: 0.04% of FS.

• Common mode rejection: 100dB at 50/60Hz.

• Zero drift: ±1 count max (autozero).

• Span tempco: ±50ppm/°C max.

• Voltage drop: ±0.1V max.

D13x Thermocouple Inputs

• Thermocouple types: J, K, T, E, R, S, B, C (factory set).

• Ranges:

J = -200°C to +760°C

B = 0°C to +1820°C

K = -150°C to +1250°C

S = 0°C to +1750°C

T = -200°C to +400°C

R = 0°C to +1750°C

E = -100°C to +1000°C

C = 0°C to +2315°C

• Resolution: ±1°.

• Overall Accuracy (error from all sources) from 0 to +40°C

ambient: ±1.0 °C max (J, K, T, E).

±2.5 °C max (R, S, B, C)(300°C to FS).

• Common mode rejection: 100dB at 50/60Hz.

• Input impedance: 100M

min.

• Lead resistance effect: <20µV per 350

.

• Open thermocouple indication.

• Input burnout protection to 250VAC.

• Overrange indication.