beautypg.com

Dimensions packing list, Ordering information specifications – Impulse MMD-4300C User Manual

Page 2

background image

Dimensions

Packing List

PACSmate® Series medical LCD monitor
Monitor stand
Power Cord (US x 1, EU x 1)
Power Adapter
Dual-link DVI-D Signal Cable
USB Cable
Utility CD-ROM

(for user's manual and software utility)

Ordering information

Specifications

PACSmate®

MMD-4300C

Drive System

a-Si TFT active matrix

Display area (H x V)

641.3mm x 400.8 mm

Diagonal size of display

75.6cm (30.0 inches)

Resolution (H x V)

2560 x 1600

Pixel pitch (H x V)

0.2505mm x 0.2505 mm

Sub-pixel pitch (H x V)

0.0835mm x 0.2505 mm

Pixel format

RGB (Red dot, Green dot, Blue dot) vertical stripe

Luminace-max.

370cd/m2

Max. Luminance (calibrated)

250cd/m2

Contrast ratio

1000:1 (typical)

Viewing angle

Right: 178°, Left: 178°, Up: 178° , Down: 178°

Response time (Ton + Toff)

12ms (typical)

Polarizer surface/ hardness

Antiglare/ 3H

Backlight

16 CCFL

Default gamma

Gamma 2.2

LUT

10-bit for each RGB

Signal input

Master: Dual link DVI-D (digital)

Slave: Single link DVI-D (digital)

USB 2.0 hub

1 upstream/ 2 downstream

Membrane keypad

Yes

Power adapter

Input voltage range

100VAC-240VAC

Power adapter

Input frequency range

47Hz-63Hz

Power adapter

Input AC current

2.0A max. at 90VAC

Power adapter

Power output

24VDC, 6.25A, 150W (max)

Physical dimensions

(H x W x D)

With stand: 522mmx 736mm x 266 mm (landscape)

Without stand: 480mm x 736mm x 76 mm (landscape)

Weight

15kg

Operating temperature

10°C~35°C

Storage temperature

0°C~45°C

Operation relative humidity

30% ~75% Non-condensing

Storage/ Transportation

Relative humidity

10% ~95% Non-condensing

DICOM

Gamma 2.2

Gamma 2.0 Gamma 1.8

User Selectable Gamma Settings and Color Temperature

Settings

PACSmate provides you with flexible Gamma settings and color temperature
settings for side-by-side image comparison in dual-viewing modes. The
independent Gamma settings (for Gamma 1.8,2.0,2.2) and color temperature
settings (6500K, 9300K) consist of multiple user-selectable settings,
providing flexible display modes for images from different modalities.

Color Temperature Settings

Built-in Front Sensor for DICOM Conformance Testing

PACSmate has designed a built-in, front-mounted sensor which continually
adjusts the luminance output to ensure consistent images over time.

Auto-Dimming for Ambient Light Adjustment

The auto-dimming function helps create better diagnostic experience under
proper light condition and extend backlight lifetime

IPX2 Compliance

The whole enclosure of MMD-4300C is compliant with IPX2 waterproof
standard.The MMD-4300C can be used in the operating room where is
humid and waterproof is considered important.

Built-in Front Sensor

Well-Lit Room

Dim-Lit Room

Part No.

Description

MMD-4300C

4 megapixel, 30" color LCD monitor with IPX2 housing

MMD-4300CX 4 megapixel, 30" color LCD monitor with IPXX housing

4MP

4MP