Standard operations – Hach-Lange BIOGAS User Manual User Manual
Page 35
35
Standard Operations
2.
Press 1 (New electrode).
3.
Press the validation key in the Function field
4.
Select the function of the electrode, refer to the
.
5.
Press the validation key in the ID field.
6.
Select Other in the From field.
Note: The option From = Catalogue allows you
to create an electrode from a list of
preprogrammed electrodes.
7.
Enter the identification (ID) of the electrode, (up
to 16 characters).
8.
Press 1 twice.
9.
Select the electrode Type, for electrodes with
pH or mV functions:
•
Single pH - pH electrode (no reference
part),
•
Combined pH - pH electrode with a refe-
rence part (with or without temperature
sensor),
•
Single metal/redox - mV (i=0) function
electrode with a reference part,
•
Combined metal/redox - mV (i=0) func-
tion electrode with a reference part (with or
without temperature sensor).
Table 3 Electrode functions
Electrode type
Select function
pH single
pH
pH combined w/wo temp. sensor pH
pH
Single metal/redox
mV (i=0)
Combined metal/redox w/wo temp. sensor
mV (i=0)
ISE single
mV (i=0)
ISE Combined w/wo temp. sensor
mV (i=0)
Reference single
Reference
Temperature sensor
T°C
Ground metal
Ground
Double metal
mV (i>0)