10 standard addition - monitoring/checking results, Standard programs – Hach-Lange DR 6000 User Manual User Manual
Page 62

62
Standard programs
Select Carousel 1 inch square to perform a series of 
measurements in 1-inch square cells with an optional 
carousel insert. 
Select Carousel 1 cm square to perform a series of 
measurements in 1-cm square cells with an optional carousel 
insert. 
Additional information can be found in the Carousel user 
manual.
5.4.10 Standard Addition - monitoring/checking results
The accuracy of readings (their correspondence with the actual 
concentration of the analyte in the sample) and their precision 
(correspondence of the reading results obtained from several 
samples containing the same concentration of the test analyte) 
can be determined or improved using the standard addition 
method
This method (also known as standard-addition or spiking) serves 
to detect sample-specific faults: 
•
The analysis was compromised by substances or 
interferences in the sample (sample matrix effect), 
•
defective measuring device,
•
impure reagents.
Method:
Add a defined quantity (concentration) of the substance to be 
tested to the sample as a standard solution. The detection rate 
(concentration of the added standard solution) should be close to 
100 %.
Detection rate
=
Reading of successful standard addition
Expanded value after successful standard addition
Detection rate
Conclusion
100 %
Probability that the reading results are correct 
is high.
≠ 100 %
Assumption: The analysis was compromised 
by substances or interferences in the sample 
(sample matrix effect)
Test to determine whether a sample matrix 
effect is present:
Use distilled water instead of the sample. Add 
standard solution as described in the 
procedure.
Detection rate
Conclusion
100 %
Ions in the sample are 
interfering with the 
analysis, causing false 
results to be obtained.
≠ 100 %
No interference ions.
