1 set up single wavelength mode, Expanded programs – Hach-Lange DR 6000 User Manual User Manual
Page 100

100
Expanded programs
4. Press OK to confirm.
Note: If a Favorite Program is deleted, it will stay in the User 
Programs or Stored Programs.
Note: If the stored program is deleted in User Programs, it will also 
be deleted in Favorites Programs.
6.3
Single Wavelength (absorbance, concentration and transmittance 
readings)
The Single Wavelength mode can be used in three ways. For 
sample readings at a single wavelength, the instrument can be 
programmed to measure the absorbance, % transmittance or 
concentration of the analyte.
•
Absorbance readings: The light absorbed by the sample is 
measured in absorbance units.
•
% Transmittance measures the percent of the original light 
that passes through the sample and reaches the detector.
•
Turning on the concentration factor allows selection of a 
specific multiplier for converting absorbance readings to 
concentration. In a graph of concentration versus the 
absorbance, the concentration factor is the slope of the line.
6.3.1
Set up Single Wavelength mode
1. Press Single Wavelength in the Main Menu.
2. Press Options for Parameter Setup.
Table 12 Options for Single Wavelength
Options
Description
More
For further Options
Recall measurement 
data symbol
Recall saved measurement data, wavelength scans or time courses, refer to
% Trans/Abs
Switches from % transmittance to concentration or absorbance in reading mode.
λ
Wavelength
To input the wavelength setting. Use the alphanumeric keypad to enter the reading wavelengths. 
A wavelength in the range of 190–1100 nm can be entered.
Timer icon
This functions as a stopwatch. It helps to make sure that the steps of an analysis are correctly 
timed (e. g. reaction times, wait times, etc., can be exactly specified). When the specified time 
has elapsed, a sound is emitted. The use of the timer has no influence on the reading program.