4 aqa - analytical quality assurance, 1 aqa standard configuration, Expanded programs – Hach-Lange DR 6000 User Manual User Manual
Page 123

123
Expanded programs
1. Press Drift Check in the "Optical Checks" menu.
2. Press
λ to select the wavelength.
3. Enter the wavelength and press OK to confirm your
selection.
4. Press Start.
The drift check lasts one hour.
This test takes a reading every minute for one hour. Every 
fifteen minutes, linear regression is used to calculate the 
slope (rate of change) for the previous fifteen minute interval.
The last reading and slope for each 15 minute interval are 
displayed. At the end of the hour, the overall values are 
calculated and displayed. The test proceeds to completion.
5. Press Cancel to return to Optical Checks.
6.7.4
AQA - Analytical Quality Assurance
1. Press AQA in the System Checks menu to configure AQA
measurement or AQA in the toolbar to the right of the 
display, to carry out AQA measurement.
The AQA menu in System Checks contains the program for 
checking the analytical quality. Configuration is done here, 
for example the individual working method being controlled by 
standard solutions or the cross-sensitivities of the analyses 
identified by standard additon solutions, multiple determinations 
and dilutions.
 As optional accessories for full individual control, standard and 
top-up solutions are available (
).
6.7.4.1
AQA Standard configuration
The working method, photometers, and accessories can be 
checked by the analysis of a standard solution.
1. Press AQA Standard
A screen with standard sets opens (not available on all 
models), alternatively an individual standard can be defined.
2. Select Options > New to define a new standard.
Note: If a standard has not yet been defined, in some models the 
alphanumeric keypad is displayed directly.
3. Enter the name of the standard solution and confirm with
OK.