Hach-Lange FP 360 sc Basic User Manual User Manual
Page 12

English 12
SENSOR SETUP
SELECT SENSOR (if there is more than one sensor)
CALIBRATE
SET OUTMODE
Behavior of the outputs during calibration and zero
point adjustment
HOLD
ACTIVE
SET TRANSFER
SENSOR MEASURE
Current, uncorrected measurement value
CONFIGURE
FACTOR: 0.1 to 100
OFFSET: –1000 to +1000
FACTOR
OFFSET
2 POINT
3 POINT
4 POINT
5 POINT
FACTOR
Is shown when FACTOR has been selected under
CONFIGURE. Refer to
for
detailed information.
OFFSET
Is shown when OFFSET has been selected under
CONFIGURE. Refer to
for
detailed information.
2-POINT
Is shown when 2 POINT has been selected under
CONFIGURE. Refer to
for
detailed information.
3-POINT
Is shown when 3 POINT has been selected under
CONFIGURE. Refer to
for
detailed information.
4-POINT
Is shown when 4 POINT has been selected under
CONFIGURE. Refer to
for
detailed information.
CALIBRATE
5-POINT
Is shown when 5 POINT has been selected under
CONFIGURE. Refer to
for
detailed information.
SET CAL DEFLT
Security query, reset to (FACTOR=1, OFFSET=0)
CONFIGURE
EDIT NAME
Name can include up to 16 characters
DEFAULT CONFIG: sensor serial number
SET PARAMETER
PAH: Measurement value related to PAH calibration
standard
OIL: Measurement value related to oil calibration
standard
DEFAULT CONFIG: PAH
SELECT UNITS
ppb, ppm, µg/L, mg/L,
DEFAULT CONFIG: ppb
AVERAGE
1 to 300 s,
DEFAULT CONFIG: 3 s
LOG SETUP
5 s, 30 s, 1, 2, 3, 4, 5, 6, 10, 15, 30 min,
DEFAULT CONFIG: 10 min
GAIN VALUE
Range 0 to 500 ppb: AUTO, 0.01 to 50, 0.01 to 500,
Range 0 to 5000 ppb: AUTO, 0.01 to 500, 0.01 to
5000
DEFAULT CONFIG: AUTO
SET DEFAULTS
Security query,
reset to default configuration for all menu options
listed above.
SENSOR SETUP (Continued)
SELECT SENSOR (if there is more than one sensor)