Appendix a modbus register – Hach-Lange FP 360 sc User Manual User Manual
Page 33

33
Appendix A Modbus register
Table 1 Sensor Modbus registers
Register#
Data type
Length
R/W
Description
40001
Unsigned integer
1
R
Reserved
40002
Float
2
R
Measurement value as PAH, unit PPB
40004
Float
2
R
Measurement value as PAH, unit PPM
40006
Float
2
R
Measurement value as OIL, unit PPB
40008
Float
2
R
Measurement value as OIL, unit PPM
40016
String
6
R
Serial number
40022
String
8
R/W
Name of location
40030
Unsigned integer
1
R/W
Choose the parameter, 47=PAH, 48=OIL
40031
Unsigned integer
1
R/W
Measurement units
38=ppb, 39= g/L, 2=ppm, 0=mg/L
40040
Float
2
R/W
Offset for calibration
40042
Float
2
R/W
Factor for calibration
40050
Unsigned integer
1
R/W
Logging interval from 5 sec to 1800 sec
40051
Unsigned integer
1
R/W
Measuring interval from 1 sec to 300 sec
40052
Unsigned integer
1
R/W
Amplification for the low/high range probe,
0=AUTO,
1= 0.01 to 50/500 PPB,
2= 0.01 to 500/5000 PPB
40055
Unsigned integer
1
R/W
Set output mode for calibration
40056
Unsigned integer
1
R/W
Set output mode for service
40057
Float
2
R
Version of application file
40059
Float
2
R
The entry is for the application file
40061
Float
2
R
Version of the probe
40063
Unsigned integer
1
R
The entry is for the device driver file
40064
Unsigned integer
1
R
The entry is for the device driver file
40065
Unsigned integer
1
R
The entry is for the device driver file
40066
Integer
1
R
Days left until exchanging lamp,
negative values show that exchange is overdue
40067
Integer
1
R
Days left until maintenance,
negative values show that service is overdue
40068
Unsigned integer
2
R
Operating hours of analyzer