Operation, 2 sensor setup (continued) – Hach-Lange FP 360 sc User Manual User Manual
Page 19

19
Operation
CONFIGURE
EDIT NAME
Name can include up to 16 characters,
DEFAULT CONFIG: sensor serial number
SET PARAMETER
PAH: Measurement value related to PAH calibration standard
OIL: Measurement value related to oil calibration standard
DEFAULT CONFIG: PAH
SELECT UNITS
ppb, ppm, μg/L, mg/L,
DEFAULT CONFIG: ppb
AVERAGE
1 to 300 s,
DEFAULT CONFIG: 3 s
LOG SETUP
5 s, 30 s, 1 min, 2 min, 3 min, 4 min, 5 min, 6 min, 10 min, 15 min, 30 min,
DEFAULT CONFIG: 10 min
GAIN VALUE
Range 0 to 500 ppb: AUTO, 0.01 to 50, 0.01 to 500,
Range 0 to 5000 ppb: AUTO, 0.01 to 500, 0.01 to 5000
DEFAULT CONFIG: AUTO
SET DEFAULTS
Security query,
reset to preset configuration for all menu options listed above.
DIAG/TEST
PROBE INFO
SENSOR NAME
Device name
EDIT NAME
SERIAL NUMBER
Device serial number
RANGE
0 to 500 or 0 to 5000
MODEL NUMBER
Item no. Sensor
CODE VERSION
Sensor software
COUNTER
OPERATING HOURS
Operating hours counter
MAINTENANCE
Counter counting down days
BULB CHANGE
Counter counting down days
TEST/MAINT
SET OUTMODE
Equipment output behavior in the SERVICE menu
HOLD
ACTIVE
SET TRANSFER
SIGNALS
LAMP CURR
Flash lamp intensity
DIAG/TEST
Zero point and slope check with external standards
REDADING OFFSET
CUBE CAL
Password-protected access for the service
4.4.2 SENSOR SETUP (Continued)
SELECT SENSOR (if there is more than one sensor)