Test the inductive conductivity sensor

If there is a doubt concerning the operation of the sensor, disconnect it from the controller but leave it
in place in your sample process. Check the following resistance values across the terminals indicated
(refer also to the wiring table

Table 1

on page 6):

Wiring connections

Resistance

Terminals 7-10 (Temperature sensor Pt100)

≈ 110 Ω (at 25 °C)

Terminals 1-2 (Secondary coil)

≈ 1—2 Ω

Terminals 11-12 (Primary coil)

≈ 1—2 Ω

If the checks do not generate any erroneous resistance values, the sensor can be considered to be
working properly.
If the measurement is still not satisfactory proceed as follows:

1. Set T-COMPENSATION to NONE (see

Inductive conductivity sensor configuration

on page 10).

2. Perform a ZERO CALIBRATION of the sensor (see

Zero calibration procedure

on page 13).

3. Perform a RESISTOR CALIBRATION of the sensor (see

Resistor calibration (inductive

conductivity sensors)

on page 16).

4. Check the following:

Displayed conductivity (K = 2.35)

Simulated resistance

470 μS/cm

5 kΩ

4.7 mS/cm

500 Ω

47 mS/cm

50 Ω

470 mS/cm

5 Ω

Sensor diagnostic and test menu

The sensor diagnostic and test menu shows current and historical information about the instrument.
Refer to

Table 2

.

To access the sensor diagnostic and test menu, push the menu key and select SENSOR
SETUP>[Select Sensor]>DIAG/TEST.

Table 2 Sensor DIAG/TEST menu

Option

Description

MODULE INFORMATION

Shows information about the sensor module.

SENSOR INFORMATION

Shows the name and serial number that was entered
by the user.

CAL DAYS

Shows the number of days since the last calibration.

CAL HISTORY

Shows a list of all calibrations by date/time stamp. Use
the arrows keys to select a calibration and push enter
to view the details.

RESET CAL HISTORY

Resets the calibration history for the sensor (requires
service-level passcode). All previous calibration data is
lost.

POLARIZATION

Contacting conductivity sensors only. Shows
information about the electrode polarization, the cable
capacitance and the time before the next
measurement.

English 19