Hach-Lange SI794 P User Manual User Manual
Page 99

99
Max. pulse frequency * 0001 ... 0180 min-1
(pulse frequency controller)
Display
LC display, 7-segment with icons
Main display
Character height 17 mm, unit symbols 10 mm
Secondary display
Character height 10 mm, unit symbols 7 mm
Sensoface
3 status indicators (friendly, neutral, sad face)
Mode indication
5 mode indicators “meas”, “cal”, “alarm”, “cleaning”, “config”
18 further icons for configuration and messages
Alarm indication
Red LED in case of alarm or HOLD, user defined
Keypad
5 keys: [cal] [conf] [
] [
] [enter]
Service functions
Current source
Current specifiable for output 1 and 2 (00.00 to 22.00 mA)
Manual controller
Controller output entered directly (start of control process)
Device self-test
Automatic memory test (RAM, FLASH, EEPROM)
Display test
Display of all segments
Last error
Display of last error occurred
Sensor monitor
Display of direct, uncorrected sensor signal
(electrode)
Relay test
Manual control of the four relay contacts
Parameter sets*
Two selectable parameter sets for different applications
Data retention
Parameters and calibration data > 10 years (EEPROM)
Protection against electric shock
Safe electrical isolation of all extra-low-voltage circuits
against mains by double insulation acc. to EN 61010-1
Power supply
24 (-15%) ... 230 V AC/DC (+10%); approx. 5 VA, 2.5 W
AC: 45 ... 65 Hz
Overvoltage category II, protection class II