beautypg.com

Orp calibration – Hach-Lange SI794 P User Manual User Manual

Page 78

background image

78

ORP calibration

ORP calibration mode is automatically preset when ORP measurement is

configured. The potential of a redox electrode is calibrated using a redox

(ORP) buffer solution. In the course of that, the difference between the

measured potential and the potential of the calibration solution is deter-

mined according to the following equation. During measurement this

difference is added to the measured potential.

0

10

20

25

30

40

50

60

70

80

Temperature

[°C]

Ag/AgCl/KCl

1 mol/L

[∆mV]

Ag/AgCl/KCl

3 mol/L

[∆mV]

Thalamid

[∆mV]

Mercury

sulfate

[∆mV]

249

244

240

236

233

227

221

214

207

200

224

217

211

207

203

196

188

180

172

163

-559

-564

-569

-571

-574

-580

-585

-592

-598

-605

672

664

655

651

647

639

631

623

613

603

Temperature dependence of commonly used reference

systems

The electrode potential can also be related to another reference system –

e.g. the standard hydrogen electrode. In that case the temperature-

corrected potential (see table) of the reference electrode used must be

entered during calibration. During measurement, this value is then added

to the ORP measured.

Please make sure that measurement and calibration temperature are the

same, since the temperature response of the reference electrode is not

automatically taken into account.

mVORP = mVmeas + ∆mV

mV

ORP

=

mV

meas

=

∆mV

=

displayed ORP

direct electrode potential

delta value, determined during

calibration