Default settings of parameter sets – Hach-Lange SI794 P User Manual User Manual
Page 63

63
Two complete parameter sets are stored in the EEPROM.
As delivered, the two sets are identical but can be edited.
Note:
Fill in your configuration data on the following pages.
Code. Parameter
Default setting
o1. pH/ORP unit
pH
o1. Electrode type
GLASS
o1. 0/4-20 mA
4-20 mA
o1. Current start
00.00 pH
o1. Current end
14.00 pH
o1. Filter time
0 s
o1. 22mA signal
OFF
o1. HOLD response
Last
o1. Fix current
021.0 mA
o2. Unit °C / °F
°C
o2. Temp probe
1000 PT
o2. 0/4 ...20mA
4-20 mA
o2. Current start
000.0 °C
o2. Current end
100.0 °C
o2. Filter time
0 s
o2. 22mA signal
OFF
o2. HOLD response
Last
o2. Fix current
021.0 mA
tc. TC measurement
Auto
tc. Measuring temp
025.0 °C
tc. TC calibration
Auto
tc. Calibration temp
025.0 °C
tc. TC medium
00.00 %/K
CA. Cal solution
-06-BUF
CA. Cal interval
0000 h
AL. Sensocheck
OFF
AL. Alarm delay
0010 sec
AL. LED Hold
OFF
Code. Parameter
Default setting
rL. Relay function
Limit
L1. Contact function
Lo
L1. Contact response
N/C
L1. Setpoint
00.00 pH
L1. Hysteresis
00.50 pH
L1. Delay
0010 sec
L2. Contact function
Hi
L2. Contact response
N/C
L2. Setpoint
14.00 pH
L2. Hysteresis
00.50 pH
L2. Delay
0010 sec
Ct. Setpoint
07.00 pH
Ct. Neutral zone
01.00 pH
Ct. P action
0100 %
Ct. I action
0000 sec
Ct. D action
0000 sec
Ct. PLC/PFC controller PLC
Ct. Pulse length
0010 sec
Ct. Pulse frequency
0060 /min
Ct. Hold response
Last
Cn. Rinse/Par.set 2
Rinse
Cn. Rinsing interval
000.0 h
Cn. Rinse duration
0060 sec
Cn. Contact type
N/C
Default settings of parameter sets