Menu

Code

Choices Default

ALrt

Alarm settings

AL.

Select Sensocheck

ON / OFF

OFF

Enter alarm delay

xxxx sec

0010 sec

LED in HOLD mode

ON / OFF

OFF

rLAY

Relay 1/2: Limit values, controller

rL.

Select limit function / controller

LiMIT / CtROL

LiMIT

Select contact function

Lo / Hi

Lo

L1.

Select contact response

N/O / N/C

N/O

Enter setpoint

xxxx

00.00 pH

Enter hysteresis

xxxx

00.50 pH

Enter delay

xxxx SEC

0010 SEC

Select contact function

Lo / Hi

Hi

L2.

Enter controller setpoint

xxxx

07.00 pH

Ct.

Enter neutral zone

xxxx

01.00 pH

(P) Controller gain K

P

xxxx %

0100 %

(I) Reset time T

R

xxxx SEC

0000 SEC

(D) Rate time T

D

xxxx SEC

0000 SEC

Pulse length/frequency controller

PLC / PFC

PLC

PLC: Pulse length

xxxx SEC

0010 SEC

PFC: Pulse frequency

xxxx /min

0060 /min

Select HOLD response

Y Last / Y Off

Y Last

Cln.

Cleaning contact

Cn.

Select cleaning contact or signaling

of parameter set 2

rinse / Pset2

rinse

rinse

Cleaning interval

xxx.x h

000.0 h

Cleaning duration

xxxx SEC

0060 SEC

Select contact response

N/O / N/C

N/O

Select contact response

N/O / N/C

N/O

Enter setpoint

xxxx

14.00 pH

Enter hysteresis

xxxx

00.50 pH

Enter delay

xxxx SEC

0010 sec