Table 3, Installation – Hach-Lange SC 1000 Controller Enhanced Communications User Manual
Page 36

36
Installation
Table 3 Status register
Bit
Status 1
Description
0
Calibration activated
Calibration in progress, measurement value not up to
date
1
Cleaning activated
Cleaning in progress, measurement value not up to
date
2
Service mode activated
Device in "Service" mode, measurement value not up to
date
3
General error message
General error detected, refer to error text for details
4
Measurement value channel 0,
poor quality
Measurement accuracy is not within specified limits
5
Measurement value channel 0,
range short-fall
Measurement value falls short of the specified range
6
Measurement value channel 0,
range exceeded
Measurement value exceeds the specified range
7
Measurement value channel 1,
poor quality
Measurement accuracy is not within specified limits
8
Measurement value channel 1,
range short-fall
Measurement value falls short of the specified range
9
Measurement value channel 1,
range exceeded
Measurement value exceeds the specified range
10
Measurement value channel 2,
poor quality
Measurement accuracy is not within specified limits
11
Measurement value channel 2,
range short-fall
Measurement value falls short of the specified range
12
Measurement value channel 2,
range exceeded
Measurement value exceeds the specified range
13
Measurement value channel 3,
poor quality
Measurement accuracy is not within specified limits
14
Measurement value channel 3,
range short-fall
Measurement value falls short of the specified range
15
Measurement value channel 3,
range exceeded
Measurement value exceeds the specified range