9 system setup menu (continued), Operation – Hach-Lange SC 100 User Manual
Page 39
Operation
System Setup Menu
35
NETWORK SETUP (this menu appears only if a network card is installed in the controller)
SET MODE
Auto or Manual
MODBUS ADDRESS
Highlight sc100 Analyzer, or either connected sensor then press
ENTER
. Choose a number between 1 and 247 as the address
(each source must have a different address) then press
ENTER
.
BAUD RATE
Select a baud rate of 9600, 19200, 38,4 K, 57,6 K, or 115,2 K. Default: 19200
STOP BITS
Select 1 or 2 stop bits. Default: 1
MODBUS MODE
Select RTU or ASCI. Default: RTU
DATA ORDER
Select Normal or Swapped. Default: Swapped
DISPLAY SETUP
ADJUST CONTRAST
Use the
UP
and
DOWN
keys to increase or decrease the contrast, see
section 4.3.1 on page 27
.
LANGUAGE
The default is English. Choose Spanish, German or French to allow all menus to appear in the selected language.
SET DATE/TIME
Use this menu to select the date format and to set the date and time (24-hour (military) format), see
section 4.3.3 on page 28
.
SECURITY SETUP (Enter a 6-digit passcode)
SET PASSCODE
DISABLE
Disables system security. See
section 4.4 on page 29
.
ENABLE
Enables system security. See
section 4.4 on page 29
.
LOG SETUP
DATALOG SETUP
Setup datalogging of displayed sensor measurements
CALCULATION
SET VARIABLE X
Select the sensor corresponding to the variable set as “X”.
SET VARIABLE Y
Select the sensor corresponding to the variable set as “Y”.
SET PARAMETER
Select the parameter to be associated with the variable.
SET FORMULA
Select the formula of the calculation to be performed on “X” and “Y”.
4.9 System Setup Menu (continued)