Installation – Hach-Lange SC 100 User Manual
Page 19
Installation
Electrical Installation
15
The DC power source that supplies power to the 24 V DC powered sc100 must maintain voltage
regulation within the specified 24 V DC –15 % / +20 % voltage limits. The DC power source must
also provide adequate protection against line transients.
Figure 12
sc100 Wiring Connections
Table 3 DC Power Wiring Information (sc100 24 V DC powered model only)
Terminal Number
Terminal Description Wire Color Code for North America
Wire Color Code for Europe
1
+24 V dc
Red
Red
2
24 V dc return
Black
Black
3
Protective Earth (PE)
Green
Green w/yellow tracer
1.
J1—Network connector
8.
Sensor connector
2.
J2—Header for optional network interface card
9.
Sensor connector
3.
J5—Relay A connector
10. J6—Analog output (4–20 mA) connector
4.
J6—Relay B connector
11. J5—Sensor connector for hard-wiring
5.
J7—Relay C connector
12. Position for network interface card
6.
Fuses (F1, F2)
13. Service port
7.
J8—Power connections
14. Sensor terminator selector/service port configuration
a.
AC Power connection (AC powered sc100 model only)
b. DC Power connection (24 V DC sc100 model only)
NC
NC
NC
COM
COM
COM
NO
F1
F2
NO
NO
RELAY C
RELAY B
RELAY A
RELAY A
J1
J2
J4
NETWORK
INTERFACE
CARD
J3
J5
J6
U5
U9
S1
6
7
7a
7b
14
12
13
3
1
2
8
9
4
5
11
10
+24
VDC
+24
VDC
–24
VDC
–24
VDC