4 setting up system security, Operation, 1 setting the passcode – Hach-Lange SC 100 User Manual
Page 33: 2 editing the passcode
Operation
Setting up System Security
29
4.4 Setting up System Security
The passcode feature of the sc100 restricts unauthorized access to configuration and calibration
settings. The passcode is factory set to
sc100 (the five digits must be followed by a space to
remove the trailing asterisk). The following two options are available:
Disabled: All configuration settings and calibrations can be changed. This is the default setting.
Enabled: Certain calibration and Test/Maint settings/operations cannot be accessed without
the passcode. If the passcode is enabled, it may be edited (see
section 4.4.2
). The passcode can
consist of up to six digits (alpha and/or numeric and available characters). If the instrument is reset
using the Configure/Default Setup menu selection, the passcode will return to the factory default. If
a changed passcode is forgotten, obtain the Master passcode from the Technical Consulting
Services Department.
4.4.1 Setting the Passcode
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
SYSTEM SETUP
3
SECURITY SETUP
4
—
SET PASSCODE
5
Highlight ENABLE
6
Main Menu or Main Measurement Screen
—
4.4.2 Editing the Passcode
Step
Select
Menu Level/Instructions
Confirm
1
MAIN MENU
—
2
SYSTEM SETUP
3
Highlight SECURITY SETUP
4
ENTER PASSCODE
5
EDIT PASSCODE
6
EDIT PASSCODE select character (shown in brackets)
—
EDIT PASSCODE move to the next character
7
Main Menu or Main Measurement Screen
—