Figure 11 Connections on the main circuit board

1 Dual monitor

connection

4 Digital inputs

7 Power out

2 Smart probe

connection

5 Power in

8 Power out LED (on =

power is connected to
the analyzer)

3 4–20 mA outputs

6 Power switch and

LED (on = analyzer is
on)

9 Relays

Connect to power

W A R N I N G

Electrocution hazard. Use a crimp-on ring terminal on the main
protective earth connection.

W A R N I N G

Electrical shock and fire hazards. Make sure that the supplied cord and non-
locking plug meet the applicable country code requirements.

W A R N I N G

Electrocution hazard. Make sure that the protective earth conductor
has a low impedance connection of less than 0.1 ohm. The connected
wire conductor must have the same current rating as the AC mains line
conductor.

N O T I C E

The instrument is used for a single phase connection only.

Cord installation: The manufacturer recommends to use the optional
cord and sealing gland. Refer to the maintenance manual for the
replacement parts list. For a customer-supplied cord, three 18-gauge
conductors are required and the cord must be shorter than 3 meters
(10 feet). Use a sealing type strain relief to keep the environmental rating
of the instrument. Refer to

Specifications

on page 3. To connect power

to the instrument, refer to

Table 3

or

Table 4

and

Figure 12

.

English 13