beautypg.com

Remove the access plugs, Remove the access cover, Wiring connections overview – Hach-Lange HACH 5500 sc Installation User Manual

Page 12

background image

W A R N I N G

Electrical shock and fire hazards. For a cord-connected instrument,
make sure to install the instrument so that the cord can be
disconnected easily from the supply socket.

N O T I C E

Make sure that the equipment is connected to the instrument in accordance with
local, regional and national requirements.

Remove the access plugs

Install cables and conduit through the electrical access ports. Refer to

Figure 9

. Remove rubber sealing plugs by pushing them out from inside

the enclosure to unlock the seal, and then remove completely by pulling
from the outside. Remove knockouts as necessary from the electrical
access plate with a hammer and screwdriver. To keep the enclosure
rating, put a cover on all ports that are not used.

Figure 9 Electrical access ports

1 Power in (power cord only), no

ground plate. Do not use for
conduit.

3 Communication and network

modules (8x)

2 Communication and network

modules (3x)

4 Power in or out (conduit or power

cord), ground plate, communication
and network modules (8x)

Remove the access cover

Remove the access cover to connect to the wiring terminals. Refer to

Figure 10

.

Figure 10 Access cover removal

Wiring connections overview

Figure 11

shows all of the possible wiring connections. Make sure to use

the wire gauge that is specified for the connection (refer to

Specifications

on page 3).

12 English