4 maintenance menus, 1 change to 0 to 20 ma output range, Operation – Hach-Lange CL17 USER MANUAL User Manual
Page 37

35
Operation
Adjust the recorder output minimum and maximum values using the analyzer keypad:
1. Press the
MENU
key.
2. Scroll to
RECR DR
menu and press
ENTER
. The display will show
REC LO
and a mg/L
reading. Move to
REC HI
or
ON SYSTEM ALARM
.
3. Press
ENTER
to change the displayed mg/L reading for
REC LO
or
REC HI
. The digit
to be changed will flash.
4. Change the digit or move to another digit. Appropriate values and selections are
shown below. When the value is correct, press
ENTER
to accept.
•
REC LO—0.0 mg/L to 5.0 mg/L limits.
•
REC HI— 0.0 mg/L to 5.0 mg/L limits.
•
ON SYSTEM ALARM—Select minimum, maximum, or hold to define the recorder
output when a system alarm is initiated.
5. Press
EXIT
to return to the main menu. Press
EXIT
again to return to normal
operation.
5.2.3.1 Change to 0 to 20 mA output range
Set the lower end of the output range to 0 mA instead of 4 mA as follows:
1. Press the
MENU
key then scroll to the
SETUP
menu. Press
ENTER
.
2. Scroll to
REC ADJUST
and press
ENTER
.
Note: A large number may be displayed for the minimum value, this number refers to counts, not
milliamps. Proceed with step 4.
3. Adjust the displayed value to 0. Press
ENTER
to accept the value. Press
EXIT
twice to
return to normal operation.
5.2.4 Maintenance menus
The following selections allow the user to perform standard maintenance tasks. Use the
ARROW
keys to scroll through the menu selections, then press
ENTER
to select the
displayed menu option.
•
PRIME—This feature causes the analyzer to make 39 concurrent accelerated cycles
to prime the reagent lines. Press
ENTER
to activate. At the end of 39 cycles, the
analyzer will return to normal operation. To terminate the prime cycle before all 39
cycles are complete, press
EXIT
. The analyzer will return to normal operation.
•
CLEAN—This feature stops the analyzer in the last portion of the measurement cycle.
Sample and reagents are left in the sample cell for cleaning purposes. The analyzer
remains in the clean mode for 60 minutes. To terminate the cycle earlier, press the
EXIT
key. The analyzer will return to normal operation.
•
HOLD OUTPUTS—This feature allows the alarms to be locked out and the recorder
output to be held at its present level for maintenance purposes. Activate this feature
as follows:
a. Press
ENTER
then the
UP ARROW
key to activate for 60 minutes. The alarm LED
will be flashing.
b. To deactivate the feature, and return to normal operation, press the
MENU
key
then the
DOWN ARROW
key until
HOLD OUTPUTS
is displayed.
c. Press
ENTER
.
d. Select
OFF
and press
ENTER
again.