35

Operation

Adjust the recorder output minimum and maximum values using the analyzer keypad:

1. Press the

MENU

key.

2. Scroll to

RECR DR

menu and press

ENTER

. The display will show

REC LO

and a mg/L

reading. Move to

REC HI

or

ON SYSTEM ALARM

.

3. Press

ENTER

to change the displayed mg/L reading for

REC LO

or

REC HI

. The digit

to be changed will flash.

4. Change the digit or move to another digit. Appropriate values and selections are

shown below. When the value is correct, press

ENTER

to accept.

REC LO—0.0 mg/L to 5.0 mg/L limits.

REC HI— 0.0 mg/L to 5.0 mg/L limits.

ON SYSTEM ALARM—Select minimum, maximum, or hold to define the recorder
output when a system alarm is initiated.

5. Press

EXIT

to return to the main menu. Press

EXIT

again to return to normal

operation.

5.2.3.1 Change to 0 to 20 mA output range

Set the lower end of the output range to 0 mA instead of 4 mA as follows:

1. Press the

MENU

key then scroll to the

SETUP

menu. Press

ENTER

.

2. Scroll to

REC ADJUST

and press

ENTER

.

Note: A large number may be displayed for the minimum value, this number refers to counts, not
milliamps. Proceed with step 4.

3. Adjust the displayed value to 0. Press

ENTER

to accept the value. Press

EXIT

twice to

return to normal operation.

5.2.4 Maintenance menus

The following selections allow the user to perform standard maintenance tasks. Use the

ARROW

keys to scroll through the menu selections, then press

ENTER

to select the

displayed menu option.

PRIME—This feature causes the analyzer to make 39 concurrent accelerated cycles
to prime the reagent lines. Press

ENTER

to activate. At the end of 39 cycles, the

analyzer will return to normal operation. To terminate the prime cycle before all 39
cycles are complete, press

EXIT

. The analyzer will return to normal operation.

CLEAN—This feature stops the analyzer in the last portion of the measurement cycle.
Sample and reagents are left in the sample cell for cleaning purposes. The analyzer
remains in the clean mode for 60 minutes. To terminate the cycle earlier, press the

EXIT

key. The analyzer will return to normal operation.

HOLD OUTPUTS—This feature allows the alarms to be locked out and the recorder
output to be held at its present level for maintenance purposes. Activate this feature
as follows:

a. Press

ENTER

then the

UP ARROW

key to activate for 60 minutes. The alarm LED

will be flashing.

b. To deactivate the feature, and return to normal operation, press the

MENU

key

then the

DOWN ARROW

key until

HOLD OUTPUTS

is displayed.

c. Press

ENTER

.

d. Select

OFF

and press

ENTER

again.