SENSOR FOR ELECTROMAGNETIC FLOW METER – MANUAL
5
ESENS_FULLBORE_05.doc
OPERATING TEMPERATURES
o
Shrewdness and precautions
NO
YES
EBONITE LINING
PP LINING
PTFE LINING
Liquid Temp.
Ambient
Temp.
Liquid Temp.
Ambient
Temp.
Liquid Temp.
Ambient
Temp.
Min. Max Min. Max Min. Max Min. Max Min. Max
Min. Max
°
C 0 80 -5 60
0
60
0 60 -20 130 -10 60
° F
30
176
23
140
32
140
32
140
-4
266
14
140
LONG PIPE
3 X DN
2 X DN
ANTI VIBRATION JOINTS
GASKET THICKNESS + 4mm
For installations on long
pipes, please use the
anti vibration joints
For vertical installations is
preferable an ascending flow
Before tightening the nuts,
place both the flanges of pipe
and device as near together
as possible.
During operation, the pipe
must be either completely full
of liquid, or completely empty
Install the sensor at a min.
distance to bows in pipes and
hydraulic accessories
For vertical installations with
descending flow direction
contact the manufacturer
Avoid operating when the
pipe is partially empty
Avoid the installation next to
bows in pipes and respect
min. inlet and outlet sections
Avoid the approach of the
flange and counter flange
using the closing force of
the nuts
Avoid the installation of the
sensor in a long pipe,
without any support of the
same
LONG PIPE