Fitting type, Code, Dim a – Brooks Instrument CMC Series User Manual
Page 9: 2 electrical installation, Function, Figure 3: d-sub 9-pin male connector pinout, 0 _installation
p.3
Figure 2: CMC Series 5-Pin Terminal Strip Dimensions
2.2 Electrical Installation
CMC Series operates from a 12-30 VDC regulated power supply. The pin outs
for the 5-Pin terminal strip, D-Sub 9-Pin male connector, and D-Sub 9-Pin male
connector with 7 inch Pig Tail are shown in Figure 3, Figure 4 and Figure 5.
NOTE: The ground of any external power supply and readout system should be the
same as the transducer ground (chassis ground) to minimize any possible ground
loops, which may effect the performance and stability of the transducer.
NOTE: The CMC Series meets CE mark requirements and complies with EMC Directive
89/336/EEC. To ensure compliance when installed, an overall metal braided
shielded cable (and Metal shielded connectors for the D-Sub connector option)
connected to chassis ground at both ends is required.
Figure 3: D-Sub 9-Pin Male Connector Pinout
[ 1.50"]
Ш
Ш
38.10 mm
[2.42"]
52.58 mm
[2.06"]
52.37 mm
Dim A
17.53mm
[0.69"]
MATING CONNECTOR
PIN 1
Ш1.50"
[Ш38.10mm]
mm
[in]
ZERO
SPAN
Fitting Type
Code
Dim A
0.5" OD Tube
0
23.95 mm [0.94"]
KF-16 Flange
2
27.68 mm [1.09"]
KF-25 Flange
3
27.68 mm [1.09"]
8 VCR
®
Female
4
42.82 mm [1.69"]
Ш1.50"
[Ш38.10mm]
0.12"
[3.05mm]
ZERO
SPAN
PIN 1
Pin
Function
1
+ Signal Output
2, 3, 5, 6,
& 7
Not Used
4
Power Supply + VDC
8
Signal Output Common
9
Power Supply Common
2.0
_INSTALLATION