beautypg.com

Flowline LD35 DeltaSpan User Manual

Page 3

background image

Rev B 

MN301035 

3 of 16 

SPECIFICATIONS  

Step Two 

Service: 

Compatible liquids 

Wetted Materials: 

Sensor:  

316 SS 

Cable: 

Polyurethane or ETFE 

Seals:  

Fluoroelastomer 

Diaphragm:   PTFE coated FKM fluoroelastomer (LD35 series only) 
Label:  

Polyethylene polyamide 

Accuracy: 

±0.25% of full scale 

Temperature Limit:  

‐4 to 176°F (‐20 to 80°C) 

Compensated 
Temperature Limits: 

LD34 series:   0 to 158°F (‐18 to 70°C) 
LD35 series:   32 to 140°F (0 to 60°C) 

Thermal Effect:  

LD34 series:   ±0.45% FS TEB 
LD35 series:   ±0.0075%/°F (±0.0135%/°C) 

Pressure Limit:  

2X full scale 

Power Requirement:  

10 to 33 VDC 

Output Signal:   

4 to 20 mA DC, 2‐wire 

Response Time: 

< 50 ms 

Max. Loop Resistance: 

1000 ohms @ 30 VDC 

Electrical Connections: 

Wire pigtail 

Cable Length: 

‐__01/‐__11:   40’ (12.2 m) 
‐__21/‐__31:   60’ (18.3 m) 

Mounting Connection: 

Suspended below point being monitored 

Electrical Protection: 

Surge/lightning protected per EN61000‐4‐5, Class 5 

Sensor Weight: 

LD34 Series:  Sensor: 0.235 lb (0.107 kg) 
LD35 Series:  Sensor: 0.3 lb (0.136 kg) 
All: 

Cable: 0.037lb (0.009 kg) per foot 

Classification: 

General Purpose 

Compliance: 

CE 


This manual is related to the following products: