beautypg.com

Flowline LD35 DeltaSpan User Manual

Page 15

background image

Rev B 

MN301035 

15 of 16 

MAINTENANCE 

Step Ten 

After  final  installation  of  the  pressure  transmitter  and  its  companion  receiver,  no  routine  maintenance  is 
required.  A  periodic  check  of  system  calibration  is  suggested.  The  pressure  transmitters  are  not  field 
repairable  and  should  be  returned  if  repair  is  needed  (field  repair  should  not  be  attempted  and  may  void 
warranty). Be sure to include a brief description of the problem plus any relevant application notes. Contact 
customer service to receive a return goods authorization number before shipping. 

Maintenance should consist of inspection to see that the transmitter is free from debris and not coated with 
any substance, which would prevent liquid from freely entering and leaving the transmitter.  If this occurs, the 
transmitter should be cleaned.   

Testing the transmitter:   

 First, verify that the sensor is wired correctly.   
 Next, check if the power supply is providing the required power.   
 Finally confirm that the loop resistance is not exceeding the sensor’s specification.   

If transmitter is not functioning properly, isolate the transmitter from the system and wire as shown below.  
Be sure to remove the sensor from the classified area when performing this test.  Multimeter should read 4 
mA with the transmitter out of liquid. 

This manual is related to the following products: