These shifts in the accuracy curve illustrate the need for field
recalibration of new flow meters after installation in place of
service.

Frequency of Calibration

If the flow meter is used in Custody Transfer service (subject
to Weights&Measures regulations), it must obviously be re-
calibrated in accordance with local W&M regulations. In
most cases these regulations call for annual re-calibration,
though different time periods apply in some territories.

If the user is ISO9000 certified, user ISO standards will indi-
cate frequency of re-calibration for instrumentation. Those
rules should be observed.

If neither W&M regulations, nor internal standards apply, our
recommendations are:
A. Calibrate immediately after installation.
B. Re-calibrate after 15-30 days.
C. Re-calibrate after 180 & 360 days.

After the run-in calibration (B) and follow-up calibrations (C),
it is possible to evaluate degree of change under normal
operating conditions. Based on values found, and total vol-
ume being metered under normal operating conditions, de-
cide whether a 6, 12 or possibly 24 month schedule should
be adopted.

Procedures & Methods

Flow meters subject to Weights&Measures regulations, or
used in systems where the flow rate can fluctuate, should
be tested at minimum, intermediate & maximum flow rates.
In non-W&M service, a flow meter always operating at a
steady flow rate, can be tested at that flow rate only.

All calibration tests should be:



Of at least 60 second duration, to minimize the effect of
flow meter error during start-up and shut-down.



Repeated 3 times to confirm flow meter repeatability.

Flow meters can be calibrated against a prover tank, a pipe
prover, a master meter or a scale. Please consult with your
distributor and/or local regulatory agencies to determine
methods available (or required).

When X = known volume dispensed, and Y = volume shown
on the register, use this formula to calculate the correction
required:

When the register is showing less than the volume actually
dispensed, the K Factor (number of pulses/unit of volume)
must be decreased accordingly. This is programmed in the
register, or possibly corrected via the mechanical calibrator if
the SCL signal conditioner is included in the assembly.

Repeat the test 2-3 times, to confirm flow meter accuracy
and repeatability.

If calibration is now correct, enter % correction on the per-
manent flow meter record, and re-seal the flow meter.

1 . 8

Flow Meter Calibration

TS Series meters in E●● & S●● assembly have an internal
pulser. The pulse signal generated by the flow meter, can
be fed directly to most electronic registers. In such assem-
blies, flow meter calibration (or re-calibration) is through
correction of the K Factor (number of pulses/unit volume) in
the electronic register.

In some cases, an optional signal conditioner is added on
the pulse signal:

PIA-300 pulse isolator/amplifier/splitter


Provides 2 identical signals out, sinking or sourcing.



Required when signal travels more than 100’ (30 m).



Recommended if signal goes to non-FPP electronics.

SCL pulse scaler/calibrator/linearizer


If pulse frequency exceeds register capacity at maximum
flow rate. This can occur at surprisingly low flow rates
(see below).



If electronic register requires a scaled pulse value (1, 10,
100 or 1000 pulses/volume unit).



If local authorities demand a mechanical calibration de-
vice. The SCL includes two dials with setting from 00 to
99. Changing the setting by 1 unit, represents approx.
0.03% correction to the pulse signal.

Flow Meter Calibration

All TS Series meters are accuracy tested prior to shipment.
Data from accuracy testing is not supplied with flow meters,
since results achieved on our test fluid do not apply to actual
liquid of operation, unless the two liquids have identical vis-
cosity characteristics.

When calibrated on a liquid with 1 cP viscosity, if the flow
meter is not field recalibrated on actual liquid of service, we
expect that the flow meter will be:



Under by 1-2% on a liquid with 0.5 cP viscosity.



Over by 0.35% to 0.6% on a liquid with 10 cP viscosity.



Over by 0.7% to 1.2% on a liquid with 100 cP viscosity.

The accuracy curve will not shift significantly at higher vis-
cosities, even if the actual operating liquid has viscosities up
to 500,000 cP.

X - Y

X

x 100

= % correction

Register m odel:
Frequency lim it:

METER

MODEL

GPM

P P G

LPM

ppl

TS10C

40

136

150

35.9

=>

91

Hz

TS15C

60

108

230

28.5

=>

108

Hz

TS20C

100

101

380

26.7

=>

168

Hz

150

72

570

19.0

=>

180

Hz

TS30C

200

51

760

13.5

=>

170

Hz

Interm. duty

250

945

=>

213

Hz

Interm. duty

300

1135

=>

255

Hz

Frequency at 100%

of model capacity.

1667 Hz

EMR

3

Pls consult with register manufacturer for other registers

ELNC

400 Hz

LCR-II™

1500 Hz