WEIGHTS

Chiller

(Dry, reservoir holds 60 Gallons)

1050

Condenser

525

HVAC Requirements:

If the condenser is installed indoors, adequate fresh air must be brought into the space and

exhausted. If the chiller section only is installed indoors, a small amount of heat is added to the space

and should be accounted for in HVAC load calculations.

Heat rejection (BTU/HR) Air Flow (CFM)

Chiller

1,800

NA

Condenser

161,000

8500

Electrical Data

Model

FLA

Max Fuse

Min Supply Circuit Amps

1055AC-480V

26 40 30

1055AC-230V

51 70 62

Each chiller has one non-fused disconnect. The branch circuit for the chiller should be protected by

fuses no larger than listed in the above table. If more than one chiller is to be installed, each chiller

requires a dedicated, fused circuit.

REMOTE TERMINAL

The remote terminal can be installed in any indoor location. It is connected to the chiller via a 6

conductor data cable. It is terminated with phone jack style connectors. The remote terminal is wall

mounted. Refer

Network requirements:

Some chiller may be shipped with optional modem or TCP/IP connection. A phone line and number is

required for the modem, and a permanent network connection is required for TCP/IP. Your network’s

router will have to be configured so that the IP address for the chiller is exposed to outside of the

facility’s LAN.

Piping Installation:

Two chilled water loops are required for each chiller, one for the cryo-compressor and one for the

water-cooled heat exchangers.

1055AC

Loop

Chiller Connection

Design Flow

Min. Pipe Size*