Schreiber Chillers 1055 New User Manual
Page 6
WEIGHTS
Chiller
(Dry, reservoir holds 60 Gallons)
1050
Condenser
525
HVAC Requirements:
If the condenser is installed indoors, adequate fresh air must be brought into the space and
exhausted. If the chiller section only is installed indoors, a small amount of heat is added to the space
and should be accounted for in HVAC load calculations.
Heat rejection (BTU/HR) Air Flow (CFM)
Chiller
1,800
NA
Condenser
161,000
8500
Electrical Data
Model
FLA
Max Fuse
Min Supply Circuit Amps
1055AC-480V
26 40 30
1055AC-230V
51 70 62
Each chiller has one non-fused disconnect. The branch circuit for the chiller should be protected by
fuses no larger than listed in the above table. If more than one chiller is to be installed, each chiller
requires a dedicated, fused circuit.
REMOTE TERMINAL
The remote terminal can be installed in any indoor location. It is connected to the chiller via a 6
conductor data cable. It is terminated with phone jack style connectors. The remote terminal is wall
mounted. Refer
Network requirements:
Some chiller may be shipped with optional modem or TCP/IP connection. A phone line and number is
required for the modem, and a permanent network connection is required for TCP/IP. Your network’s
router will have to be configured so that the IP address for the chiller is exposed to outside of the
facility’s LAN.
Piping Installation:
Two chilled water loops are required for each chiller, one for the cryo-compressor and one for the
water-cooled heat exchangers.
1055AC
Loop
Chiller Connection
Design Flow
Min. Pipe Size*