Specifications, Dc voltage, Ac voltage – B&K Precision 5491B - Datasheet User Manual
Page 3: Dc current
2
8
3
1
E
5
4
9
1
B
3
w w w.bk prec ision.c om
True RMS Bench Multimeters
Models 2831E and 5491B
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Re solution
Fu ll Scale
Re ading
Accu racy
(1 year)
T ypical I nput
I mpedance
Slow
200.00 mV
10 µV
210.00
0.03%+0.08% (1)
>10 MΩ
2.0000 V
100 µV
2.1000
0.03%+0.05% (1)
>11.1 MΩ
20.000 V
1 mV
21.000
0.03%+0.06%
>10.1 MΩ
200.00 V
10 mV
210.00
0.03%+0.06%
10 MΩ
1000.0 V
100 mV
1010.0 (2)
0.03%+0.06%
10 MΩ
Slow
500.00 mV
10 µV
510.00
0.02%+0.016% (1)
>10 MΩ
5.0000 V
100 µV
5.1000
0.02%+0.008% (1)
>11.1 MΩ
50.000 V
1 mV
51.000
0.02%+0.008%
>10.1 MΩ
500.00 V
10 mV
510.00
0.02%+0.008%
10 MΩ
1000.0 V
100 mV
1010.0 (2)
0.02%+0.008%
10 MΩ
(1) under REL status
(2) 1% over-range (1010 V) is readable at 1000 V range
DC Voltage
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Re solution
Full Scale
Readi ng
Accuracy(1 ye ar)(1) 23 ° C ± 5 ° C
20~50 Hz
50~20 kH z
20~50 kHz
50~100 kHz
S low
200.00 mV
10 µV
210.00
1.0%+0.2%)
0.5%+0.15%
1.8% + 0.25%
3.0% + 0.75%
2.0000 V
100 µV
2.1000
1.0%+0.2%)
0.4%+0.05%
1.5% + 0.1%
3.0% + 0.25%
20.000 V
1 mV
21.000
1.0%+0.2%
0.4%+0.05%
1.5% + 0.1%
3.0% + 0.25%
200.00 V
10 mV
210.00
----
0.8%+0.075%
1.5% + 0.1%
3.0% + 0.25%
750.0 V
100 mV
757.5(3)
----
0.8%+0.075%
1.5% + 0.1% (2)
3.0% + 0.25% (2)
S low
500.00 mV
10 µV
510.00
1.0%+0.08%)
0.5%+0.06%
1.5% + 0.1%
3.0% + 0.3%
5.0000 V
100 µV
5.1000
1.0%+0.08%)
0.35%+0.02%
1% + 0.04%
3.0% + 0.1%
50.000 V
1 mV
51.000
1.0%+0.08%
0.35%+0.02%
1% + 0.04%
3.0% + 0.1%
500.00 V
10 mV
510.00
----
0.5%+0.03%
1% + 0.04%
3.0% + 0.1%
750.0 V
100 mV
757.5(3)
----
0.5%+0.03%
1% + 0.04% (2)
3.0% + 0.1% (2)
Max. crest factor: 3.0 at full scale
(1) Specifications are for sine wave inputs >5% of range.
(2) Limit at 40 kHz or ≤ 3×107 Volt-Hz for 750 V range
(3) 1% over-range (757.50V) is readable at 750V range
AC Voltage
Specifications
2
8
3
1
E
5
4
9
1
B
2
8
3
1
E
5
4
9
1
B
Resolution, Full Scale Reading and Accuracy ± (% of reading + % of range), 23 °C ± 5 °C
Rate
Range
Resolu tion
Fu ll Scale
Re ading
Accuracy
(1 ye ar)
B urden Voltage (1)
& Shu nt Re sistor
S low
2.0000 mA
0.1 µA
2.1000
0.08%+0.025% (2)
<0.3 V / 100 Ω
20.000 mA
1 µA
21.000
0.08%+0.02% (2)
<0.04 V / 1 Ω
200.00 mA
10 µA
210.00
0.08%+0.02%
<0.3 V / 1 Ω
2.0000 A
100 µA
2.1000
0.3%+0.025%
<0.05 V / 10 mΩ
20.000 A
1 mA
21.000 (3)
0.3%+0.025%
<0.6 V / 10 mΩ
S low
5.0000 mA
0.1 µA
5.1000
0.05%+0.01% (2)
<0.6 V / 100 Ω
50.000 mA
1 µA
51.000
0.05%+0.008% (2)
<0.06 V / 1 Ω
500.00 mA
10 µA
510.00
0.05%+0.008%
<0.6 V / 1 Ω
5.0000 A
100 µA
5.1000
0.25%+0.01%
<0.1 V / 10 mΩ
20.000 A
1 mA
21.000 (3)
0.25%+0.01%
<0.6 V / 10 mΩ
(1) Typical voltage across the input terminals at full scale reading.
(2) Use REL function
(3) In 20 A range, >10~20 ADC is readable for 20 seconds maximum
DC Current