Physical characteristics, Mechanical, Automated video measurement set — vm6000 – Atec Tektronix-VM6000 User Manual
Page 21
![background image](/manuals/320054/21/background.png)
Automated Video Measurement Set — VM6000
Physical Characteristics
Benchtop Configuration
Dimension
mm
in.
Height
292
11.48
Width
451
17.75
Depth
265
10.44
Weight
kg
lb.
Net
15
32
Shipping
28.9
63.75
Rackmount Configuration
Dimension
mm
in.
Height
323
12.25
Width
479
18.85
Depth (from rackmounting
ear to back of instrument)
231.75
9.12
Weight
kg
lb.
Net
17.4
37.5
Kit
2.5
5.5
Mechanical
Cooling – Required Clearance
Dimension
mm
in.
Top
0
0
Bottom
0
0
Left side
0
0
Right side
76
3
Front
0
0
Rear
0
0
Environmental
Characteristic
Description
Temperature
Operating
+10 °C to +45 °C
Nonoperating
-40 °C to +71 °C
Humidity
Operating
5% to 95% relative humidity (RH) with a maximum
wet bulb temperature of +29 °C at or below +50 °C,
noncondensing. Upper limit derated to 45% RH above
+30 °C up to +50 °C
Nonoperating
5% to 95% relative humidity (RH) with a maximum
wet bulb temperature of +29 °C at or below +60 °C,
noncondensing. Upper limit derated to 45% RH above
+30 °C up to +50 °C
Altitude
Operating
10,000 ft. (3,048 m)
Nonoperating
40,000 ft. (12,190 m)
Random Vibration
0.000125 G
2
/Hz from 5 to 350 Hz
-3 dB/octave from 350 to 500 Hz
0.0000876 G
2
/Hz at 500 Hz
Operating
Overall level of 0.27 G
RMS
0.0175 G
2
/Hz from 5 to 100 Hz
-3 dB/octave from 100 to 200 Hz
0.00875 G
2
/Hz from 200 to 350 Hz
-3 dB/octave from 350 to 500 Hz
0.006132 G
2
/Hz at 500 Hz
Nonoperating
Overall level of 2.28 G
RMS
Regulatory
Electromagnetic
Compatibility
93/68/EEC; EN61326:1997 +A1 1998+A2:2000
Certifications
UL 3111-1, CSA1010.1, ISO11469,EN61010-1, IEC
61010-1