Wt500 specification, Harmonic measurement function (/g5 optional), Delta calculation function (/dt optional) – Atec Yokogawa-WT500 User Manual
Page 3: Ethernet communications (/c7 optional), Usb port (pc), Usb port (peripheral), Gp-ib interface (/c1 optional), General specifications
WT500 SPECIFICATION
Harmonic Measurement Function (/G5 Optional)
• During Harmonic Display
10 Hz
≤f<45 Hz
45 Hz
≤f≤440 Hz
440 Hz ≤1 kHz 1 kHz ≤2.5 kHz 0.4% of reading + 0.35% of range 0.75% of reading + 0.35% of range 1.2% of reading + 0.35% of range 5% of reading + 0.35% of range 0.85% of reading + 0.5% of range 1.5% of reading + 0.5% of range 2.4% of reading + 0.5% of range 10% of reading +0.5% of range • When Line Filter is ON (5.5 kHz) 10 Hz ≤f<45 Hz 45 Hz ≤f≤440 Hz 440 Hz ≤1 kHz 1 kHz ≤2.5 kHz 2.5 kHz ≤5 kHz 0.15% of reading + 0.35% of range 0.15% of reading + 0.35% of range 0.2% of reading + 0.35% of range 0.8% of reading + 0.35% of range 3% of reading + 0.35% of range 0.15% of reading + 0.35% of range 0.15% of reading + 0.35% of range 0.2% of reading + 0.35% of range 0.9% of reading + 0.35% of range 3% of reading + 0.35% of range 0.35% of reading + 0.5% of range 0.25% of reading + 0.5% of range 0.4% of reading + 0.5% of range 1.7% of reading + 0.5% of range 6% of reading + 0.5% of range • When Line Filter is OFF Fundamental Frequency 10 Hz to 75 Hz 75 Hz to 150 Hz 150 Hz to 300 Hz 300 Hz to 600 Hz 600 Hz to 1200 Hz Sample Rate f*1024 f*512 f*256 f*128 f*64 Window Width 1 2 4 8 16 Upper Limit of Analyzed orders 50 32 16 8 4 Sampling Frequency Voltage Current Power Sampling Frequency Voltage Current Power However, all the items below apply to all tables. λ (power factor) = 1 • Power figures that exceed 440 Hz are reference values. order of the voltage and current. For the n+mth order and n-mth order of power, add {n/(m+1)/25} of the nth order reading. nth component of the power. • Accuracy when the crest factor is 6: The same as when the range is doubled for crest factor 3. normal measurement. If the amplitude of the high frequency component is large, influence of approximately Delta Calculation Function (/DT Optional) FQ) is installed, the frequencies of the voltages and currents Measuring Range 100 ms 25 Hz ՅfՅ100 kHz 200 ms 12.5 Hz ՅfՅ100 kHz 500 ms 5 Hz ՅfՅ100 kHz 1 s 2.5 Hz ՅfՅ100 kHz 2 s 1.5 Hz ՅfՅ50 kHz 5 s 0.5 Hz ՅfՅ20 kHz ±0.06% of reading RGB Video Signal (VGA) Output Section (/V1 Optional) 15-pin D-Sub (receptacle) Ethernet Communications (/C7 Optional) 1 Conforms to IEEE 802.3. USB port (PC) Type B connector (receptacle) Max.12 Mbps USB port (Peripheral) Type A connector (receptacle) Conforms to USB Rev.2.0 Master/Slave Synchronization Signal Input/External Clock Input (Select) BNC connector: Both slave and master BNC connector Same as the measurement range for frequency 10 Hz to 1.2 kHz 1 µs Within (1 µs + 1 sample rate) GP-IB Interface (/C1 optional) Use one of the following by NATIONAL INSTRUMENTS: IEEE St'd 488-1978 (JIS C 1901-1987). General Specifications Approximately thirty minutes. °C 20–80% (when printer not used) °C (no condensation may be present) 20 to 80% RH (no condensation) 90–264 VAC 48 to 63 Hz All Installed Elements Item Voltage Current Delta Calculation Setting difference 3P3W→3V3A DELTA→STAR STAR→DELTA difference 3P3W→3V3A DELTA→STAR STAR→DELTA Symbols and Meanings ̅U1: Differential voltage determined by computed u1 and u2 ̅U1, ̅U2, ̅U3: Line voltage determined in the calculation for a 3 phase 4 ̅11: Differential current determined by computation Neutral line current Neutral line current Sample rate (sampling frequency), window width, and upper limit of analyzed orders for PLL Accuracy ±(reading error + measurement range error) (for crest factor 3) Measurement method Accuracy Max. display resolution Connector type Number of communication ports Transmission system Connector Power Supply Connector Speed Supported USB memory devices Master/Slave Synchronization Signals External Clock Input For Triggers Card driver Conforms electrically and mechanically Functional specification Warm-up time Operating altitude Rated supply frequency Maximum power consumption Measure source PLL source Data length for FFT
• When the crest factor is set to 3
• When
• For nth order component input, add {n/(m+1)}/50% of (the nth order reading) to the n + mth order and n-mth
• Add (n/500)% of reading to the nth component of the voltage and current, and add (n/250)% of reading to the
• The accuracy guaranteed range by frequency and voltage/current is the same as the guaranteed range of
1% may appear in certain orders. The influence depends on the size of the frequency component. Therefore, if
the frequency component is small with respect to the range rating, this does not cause a problem.
being input to all input elements can be measured.
Reciprocal method
Data Update Rate
When the input signal levels are greater than or equal to 25
mV (current external sensor input) and the signal is greater
than or equal to 30% (0.1 Hz–440 Hz, frequency filter ON), of
the measurement range.
However, when the measuring frequency is smaller or equal
to 2 times of above lower frequency, the input signal is
greater than or equal to 50%.
Add 0.05% of reading when current external input is smaller
than or equal to 50 mV input signal level for each is double for
crest factor 6.
99999
0.0001 Hz
Select ON/OFF
VGA compatible
RJ-45 connector
Ethernet 100BASE–TX
Max.100 Mbps
TCP/IP
FTP server, DHCP, DNS, Remote control (VXI-11)
1
Remote control (USB-TMC)
Models with standard USB ports that run Windows 2000,
Windows XP, or Windows Vista with USB port
as a standard.
Self Power
Max. 480 Mbps
2
104 keyboard (US) and 109 keyboard (Japanese) conforming
to USB HID Class Ver.1.1devices
USB (USB Mass Storage Class) flash memory
5 V, 500 mA (per port)
However, device whose maximum current consumption
exceeds 100 mA cannot be connected simultaneously to the
two ports.
TTL
50% duty ratio square wave
50% duty ratio square wave
• AT-GPIB
• PCI-GPIB, PCI-GPIB+, and PCIe-GPIB
• PCMCIA-GPIB and PCMCIA-GPIB+
Use driver NI-488.2M version 1.60 or later.
SH1, AH1, T6, L4, SR1, RL1, PP0, DC1, DT1, and C0.
IEEE St'd 488.2-1992.
ISO (ASCII)
Addressable mode
0–30
Remote mode can be cleared using the LOCAL key (except
during Local Lockout).
5–40
(No condensation may be present)
2000 m or less
Inside of room
-25–60
100–240 VAC
50/60 Hz
80 VA (when using built-in printer)
Approximately 6.5 kg (including main unit, 3 input elements,
and options)
PLL synchronization
PLL source of the fundamental frequency is in the range 10
Hz–1.2 kHz.
Select voltage, current, or external clock for each input
element.
32 bits
Rectangular
Set using a line filter (5.5 kHz or OFF)
̅U1: Line voltage determined in the calculation for a 3 phase 3 wire connection
̅U1, ̅U2, ̅U3: Phase voltage determined in the calculation for 3 phase
3 wire (3V3A) connection
wire connection
Phase current that are not measured can be computed
synchronization.
Measurement range
Min. frequency resolution
Frequency Filter
Output format
Connector type
Electrical and mechanical specifications
Transmission rate
Protocol
Supported Services
Electrical and Mechanical Specifications Conforms to USB Rev.1.1
Speed
Number of Ports
Supported service
Supported Systems
Electrical and Mechanical Specifications
Number of Ports
Supported keyboards
Power supply
Connector type
Connector type
Input level
Inputting the synchronization source as the Ext Clk of normal measurement.
Frequency range
Input waveform
Inputting the PLL source as the Ext Clk of harmonic measurement. (/G5 option is required)
Frequency range
Input waveform
Minimum pulse width
Trigger delay time
Conforms to protocol
Encoding
Mode
Address
Clear remote mode
Operating temperature:
Operating humidity:
Operating area
Storage environment:
Storage humidity:
Rated supply voltage
Allowed supply voltage fluctuation range
Allowed supply frequency fluctuation
Weight
Method
Frequency range
Window function
Anti-aliasing filter