High Performance Power Analysis System

2503AH

Tel: (858) 530-8099 Fax: (858) 530-8077

www.xitrontech.com Email: [email protected] [email protected]

XITRON Technologies Incorporated 9770-A Carroll Centre Road San Diego, California 92126

® Copyright 2005 XITRON Technologies Inc.

Input Ranges

User may select fixed or autorange.
Voltage: 15-30-60-150-300-600-1200 Vrms
Current: Shunt: 0.05-0.1-0.2-0.5-1-2-5-10-20 Arms
*Int. CT: 7.5-15-30-60 Arms
Bypass: 12.5-25-50-125-250-500 mV RMS, 1.25-2.5-5V RMS
All ranges allow for up to 2.5X range peak

Resolution

Better than 0.05% of range

Voltage & Current Accuracy

DC Volts: 0.05% +/- 0.15% range +/- 50 mV
DC Amp: 0.05% +/- 0.15% range +/- 200 µA

AC Volts/Amp

0.001Hz-10 kHz 0.05%

10kHz-20kHz

0.10%

20kHz-50kHz

0.33%

50kHz-100kHz

0.55%

100kHz-200kHz

1.00%

200kHz-500kHz 2.35%

For voltage add 0.05% of range + 20 mV
For internal shunt add 0.05% of range + 100 µA
For shunt bypass add 0.05% of range + 10 µV
Min input > 10% of range (1% with filter on)

High Accuracy Option

0.05% of reading for freq. 40-400 Hz, and input >25% of range

Hall Effect CT* Accuracy

DC Amp: 0.15% +/- 0. 15%, range +/- 25mA
AC Amp: 0.1Hz- 10kHz:

0.25%

10kHz-20kHz

0.65%

20kHz-50kHz:

2.25%

50kHz-100kHz:

4.25 %

For AC add 0.05% of range + 10 mA

Crest Factor

Better than 2.5 at full scale input, linearly increasing to 250:1
at 1% of full scale. For max. inputs of 100 Apk, 3000 Vpk

Voltage Protection

Up to 3000 Vpk. Max slew rate 2500 V/uSec

Current Protection

Max 500 Amp peak via HALL effect CT*

Max 15V peak using shunt bypass input

Max. 50 Amp peak using internal shunt

Isolation

Inputs are isolated from each other and ground for voltages up to 3000 Vpk

Settling Time

0.0015 mSec (low pass filter disabled)

Low Pass Filters

User definable 5 Hz - 250 kHz,or disabled

Filter Amplitude Accuracy

Add 0.01%/kHz for signal frequencies > 5 kHz, Filter rejection > 40 dB @ 3x
selected filter frequency, current and voltage accuracy specifications apply for
input signals < 0.05x selected filter frequency

Harmonic & Spectrum Analysis

Bandwidth: 0.001 Hz to 170 kHz
Max. harmonic: 2047
Max FFT size: 4096 point complex FFT, Typical THD, harmonic and phase
accuracy at line frequencies of 50/60 Hz
THD accuracy: +/- 0.3%
Harmonic accuracy: 0.03% of range
Phase accuracy: 0.1° for freq., <5 kHz, linearly increasing to
5° @ 170 kHz, Fully compliant IEC 1000-3-2 harmonic analysis (Pre-A14)

Power Factor Accuracy

Approximately 0.001 for freq. 10 kHz (5 kHz w/filter) increasing linearly to 0.01
@ 200 kHz (20 kHz w/filter)

AIO Option

Up to twelve analog outputs, proportional to user offset & span of parameters
Analog output level/resolution: 5V/5mA @ 16 bits
Accuracy: 0.1% +/- 2mV
Digital Range Output: Contact Xitron Technologies

Physical specifications

Power input: 85-265 VRms autoselect, 40-400 Hz @ 100VA max
Size:

17.71" wide by 7" high by 14" deep

Weight: 24 lbs., 35 lbs. Shipping weight
Operating range: 0°C to 50°C, <85% RH @ 40°C non-condensing
Storage range: -30°C to 65°C <95% RH @ 40°C non-condensing
Configuration: Benchtop or optional 19" rack mount

Digital interfaces (standard)

IEEE488 (1), RS-232 (2) , Parallel Printer

Options

HA: High accuracy calibration 40-400Hz, 0.05% all parameters
AIO:

12ea. analog out, 16 digital out

HE 1CH Internal Hall effect for single channel analyzer*
HE 2CH Internal Hall effect for two channel analyzer*
HE 3CH Internal Hall effect for three channel analyzer*
RE 19" Rack Adapter

* Internal HALL effect CT options not available on CE marked units

Condensed
Specifications

(Contact XITRON for complete specifications)