Condensed specifications – Atec Xitron-2553 User Manual
Page 2
Condensed Specifications
(Contact XiTRON for complete specifications)
Voltage: Range
950Vpk ,400Vpk ,1500Vpk (see ordering information)
Internal Current: Range
8Apk ,40Apk (see ordering information)
External Current Transducers (Option E only)
Current: Current
Current: 10Arms for 50ms, 2.5Arms continuous, 5Apk measurable
Scaling (In=Out) = +0000.00A = 00.0000A Through +\-9999.99A = 9999.99A
Current: Voltage
Voltage: 25Vpk for 50ms, 2.5Vrms continuous, 2.5Vpk measurable
Scaling (In=Out) = +0000.00V = 00.0000V Through +\-9999.99V = 9999.99V
Resolution
0.05% of range
Voltage & Current Accuracy
(See chart to the left)
Frequency Measurement
20 mHz to 80 kHz, 0.01%
Crest Factor
Range: 1 to 99
Accuracy: from peak results
Harmonic & Spectrum Analysis
Frequency Range: 20 mHz to 50 kHz
Typical accuracy at line frequencies of 50/60 Hz:
THD:0.1%
Harmonic: 0.05%
Phase: 0.1º
Waveforms
Actual, Peak Capture and Distortion content
History
V&A (rms, peak, envelope, THD), Watts, VAR, PF
User-selectable time base from 400 msec/div to 1 day/div
Resolution 1/20 division
Physical
Power input: 12VDC @ 1.5A minimum output
Size: (HxWxD) 4.7” x 13.8”x 9.5” (11.94cm x 35.05cm x 24.13cm)
Weight: 7.5 lbs (3.4 kg)
Operating range: 0ºC to 45ºC, <85% RH @ 40ºC non-condensing
Storage range: -30ºC to 65ºC <95% RH @ 40ºC non-condensing
Unit is supplied with one XiTRON Technologies T5 universal
external power supply, 100-240Vrms, 50-60Hz, with a 2.5mm 12VDC output plug
and a three prong IEC320 AC inlet receptacle, plus a three-prong AC power cord.
Digital Interfaces (standard)
IEEE488.1: Full talk/listen capabilities
Parallel Printer: IEEE1284, unformatted text or PCL Compatible
Warranty
Two years
9770-A Carrol Centre Road | San Diego, CA 92126 | 858.530.8099 phone | 858.530.8077 fax
www.xitrontech.com | [email protected] | [email protected]
2553
Voltage & Current Accuracy
Input Configuration
Ordering Information
Standard
or Option E
A
V
Option I
V
A
LOAD
SOURCE
CURRENT
VOLTAGE
Option
Max Volts
Max Current
-
400V
8A
E
950V
40A
-
1500V
For Example: To order a base configuration unit that has External
Current Transducer capability, can measure Voltages up to 950V
peak and can also measure up to 40 Amps peak when configured
for Internal then order as follows: 2553 - E - 950V - 40A.
Copyright
©
2007 - XiTRON Technologies, Inc.