Atec Tektronix-TDS3000 Series User Manual
Page 3
T D S 3 0 0 0 S E R I E S D I G I T A L P H O S P H P O R O S C I L L O S C O P E S
3
WAVEFORM PROCESSING
Deskew – Channel to channel deskew ±10 ns
may be manually entered for better timing mea-
surements and more accurate math waveforms.
Arithmetic Operators – Add, Subtract,
Multiply, Divide.
Autoset – Single-button, automatic setup on
selected input signal for vertical, horizontal, and
trigger systems.
DISPLAY CHARACTERISTICS
Graticules – Full, grid, cross-hair, frame, NTSC,
PAL, SECAM, vectorscope 100% and 75% color
bars (with optional TDS 3VID and TDS 3SDI video
application modules).
Format – YT, XY and Gated XYZ (XY with Z-axis
blanking available on TDS 30X4 only).
I/O INTERFACE
Hardcopy Port (standard) – Centronics-type
parallel.
TDS 3GM Communications Module –
GPIB (IEEE –488.2) Programmability: Full talk/lis-
ten modes; Control of all modes, settings, and
measurements.
RS-232-C Interface Programmability: Full talk/lis-
ten modes; Control of all modes, settings, and
measurements. Baud Rate up to 38,400. DB-9
male connector.
Programmer Manual: (071-0381-00).
TDS 3VM Communications Module –
VGA: Monitor output for direct display on large
VGA-equipped monitors. DB-15 female connector,
31.6 kHz sync rate, EIA RS-343A compliant.
RS-232-C Interface Programmability: same as
TDS 3GM.
Programmer Manual: same as TDS 3GM.
TDS 3EM Communications Module –
Ethernet Port: 10Base-T with RJ-45 connector.
Provides local area network printing and pro-
gramming interface.
RS-232-C Interface programmability: same as
TDS 3GM.
Programmer Manual: same as TDS 3GM.
Note: Only one Communication Module may be
installed at a time.
All Communication Modules include WaveStar
™
Software for oscilloscopes 30-day, full-functioning
product demo.
HARD COPY CAPABILITY
Graphics File Formats – Interleaf (.img), TIF,
PCX (PC Paintbrush), BMP (Microsoft Windows),
and Encapsulated Postscript (EPS).
Printer Formats – Bubblejet, DPU-3445,
Thinkjet, Deskjet, Laserjet, Epson (9 and 24-pin).
ENVIRONMENTAL AND SAFETY
Temperature – +5 to +50°C (operating), –20 to
+60°C (nonoperating).
Humidity – 20% to 80% RH below 32°C, derate to
30% RH at 45°C (operating), 5% to 90% RH below
41°C, derate to 30% RH at 60°C (nonoperating).
Altitude – to 3,000 m (operating), 15,000 m
(nonoperating).
Electromagnetic Compatibility – Meets or
exceeds EN55011 Class A Radiated and
Conducted Emissions; EN50082-1; FCC 47 CFR,
Part 15, Subpart B, Class A; Australian EMC
Framework; Russian GOST EMC regulations.
Safety – UL3111-1, CSA1010.1, EN61010-1,
IEC61010-1.
P h y s i c a l C h a r a c t e r i s t i c s
Instrument
Dimensions
mm
in.
Width
375.0
14.8
Height
176.0
6.9
Depth
149.0
5.9
Weight
kg
lb.
Instrument only
3.2
7.0
w/battery
5.2
11.5
Characteristics (Continued)
TDS 3012, TDS 3014, TDS 3032, TDS 3034,
TDS 3052, TDS 3054
STANDARD ACCESSORIES
Probes: 2 each P3010 10X passive probes (TDS 3012), 4
each P3010 10X passive probes (TDS 3014), 2 each
P6139A 10X passive probes (TDS 3032, TDS 3052), 4
each P6139A 10X passive probes (TDS 3034, TDS 3054).
Documentation: Quick Reference Manual, User Manual,
On-line Tour Disk, Front panel overlay for non-English
languages.
Application Modules (TDS 3014, TDS 3034, TDS 3054
only): TDS 3FFT, TDS 3TRG.
Power Cord.
Accessory Tray.
Protective Front Cover: has holder for reference manual
and/or 3.5 in. floppy disks.
NIST-Traceable Certificate of Calibration.
WARRANTY INFORMATION
Three year warranty covering all labor and parts,
excluding probes.
INTERNATIONAL POWER PLUGS
Standard – US (161-0104-00).
Opt. A1 – Universal Euro 220 V, 50 Hz (161-0104-06).
Opt. A2 – United Kingdom 240 V, 50 Hz (161-0104-07).
Opt. A3 – Australia 240 V, 50 Hz (161-0104-05).
Opt. A5 – Switzerland 220 V, 50 Hz (161-0167-00).
Opt. A6 – Japan 3 to 2 wire adapter (013-0310-00).
INTERNATIONAL USER MANUALS (TDS 3000
SERIES, TDS 3FFT, TDS 3TRG, TDS 3VID,
TDS 3TMT)
Standard – English.
Opt. L1 – French.
Opt. L2 – Italian.
Opt. L3 – German.
Opt. L4 – Spanish.
Opt. L5 – Japanese.
Opt. L6 – Portuguese.
Opt. L7 – Simplified Chinese.
Opt. L8 – Traditional Chinese.
Opt. L9 – Korean.
Opt. LR – Russian. (TDS 3000 products only)
INSTRUMENT ACCESSORIES
TDS 3SDI – 601 Digital Video Module.
Converts ITU-R BT.601 format serial digital video to ana-
log video. Automatically detects and displays 525/60
and 625/50 formats. Provides composite and compo-
nent (RGB, YPbPr, or YC) outputs. Provides auto equal-
ization for cable lengths up to 250 m and EDH and
counting for incoming SDI signal. Adds video picture
mode with on-screen line select and vectorscope mode
with 100% and 75% color bars. Adds triggering and
vectorscope graticules for analog HDTV. Includes full
functionality of TDS 3VID Extended Video Applications
module. Includes special output cable and four 75
Ω
terminators. Ships with Quick Reference and User
Manuals in 10 languages. (Russian not available.)
Ordering Information
TDS 3BAT – Battery Pack being installed.
Application Module being installed.