beautypg.com

Characteristics, Ordering information, P6248 – Atec Tektronix-P6046 User Manual

Page 2: P6247, P6246, Datasheet characteristics

background image

Datasheet

Characteristics

Characteristic

Description

Bandwidth

(Guaranteed)

P6248: DC to 1.5 GHz

P6247: DC to 1.0 GHz

P6246: DC to 400 MHz

Rise Time

(Specified)

<265 ps (P6248)

<350 ps (P6247)

<875 ps (P6246)

DC Attenuation

Accuracy

±2%

Common

Input Voltage Range

±7.0 V, 1X

±7.0 V, 10X

Differential

Input Voltage Range

±850 mV, 1X

±8.5 V, 10X

Max Nondestruct

Input Voltage

±25 V (DC + peak AC)

Electrostatic

Immunity

>15 kV

TekProbe

®

Interface Probe power, auto-scaling, auto-termination with TDS

oscilloscopes

Bandwidth Limit

Switchable, 200 MHz (P6246 and P6247)

DC Reject

0.4 Hz, 1X

4.0 Hz, 10X (P6246 and P6247)

Typical Specifications

Characteristic

Description

Input C

<1 pF (Differential mode)

<2 pF (Common mode)

Input R

200 kΩ (Differential mode)

100 kΩ (Common mode)

Linearity

±2%

Noise

<50 nV/squareroot Hz

Harmonic Distortion ≤1.5%
Passband Ripple

< ±0.25 dB

P6248: DC to 1.5 GHz

P6247: DC to 850 MHz

P6246: DC to 400 MHz

Output Termination Terminate into 50 Ω

Physical Characteristics

Characteristic

Description

Weight (Probe only) 160 g (0.36 lb.)
Probe Head Volume 6 cm

3

(0.3 in.

3

)

Probe Head Input

Connection

0.63 mm (0.025 in.) square pins on 2.54 mm (0.1 in.) centers

Output Cable

Length

1.2 m (4.0 ft.)

Environmental Characteristics

Characteristic

Description

Temperature

Operating

0 °C to 50 °C

Nonoperating

–55 °C to +75 °C

Humidity (0 to 90% RH)

Operating

30 °C to 50 °C

Nonoperating

30 °C to 60 °C

P6246/P6247/P6248 CMRR

Frequency

Attenuation

Setting

P6246

P6247

P6248

÷1

>60 dB

>60 dB

>60 dB

1 MHz

÷10

>45 dB

>45 dB

>45 dB

÷1

>38 dB

>38 dB

>38 dB

100 MHz

÷10

>25 dB

>25 dB

>25 dB

÷1

N/A

>35 dB

>35 dB

500 MHz

÷10

N/A

>20 dB

>20 dB

÷1

N/A

>30 dB

>30 dB

1 GHz

÷10

N/A

>18 dB

>18 dB

÷1

N/A

N/A

>25 dB

(typical)

1.5 GHz

÷10

N/A

N/A

>18 dB

(typical)

Ordering Information

P6248

1.7 GHz Differential Probe (typical ≤27 °C). 1.5 GHz Differential Probe (guaranteed).

Includes: Certificate of Traceable Calibration, User Manual (English):

Order 071-0566-00, Service Manual: Order 071-0573-00, Carrying Case:

Order 016-1787-00.

P6247

1.0 GHz Differential Probe (guaranteed).

Includes: Certificate of Traceable Calibration, User Manual (English, German,

French and Japanese): Order 070-9898-02, Service Manual: Order 070-9899-01,

Carrying Case: Order 016-1787-00.

P6246

400 MHz Differential Probe (guaranteed).

Includes: Certificate of Traceable Calibration, User Manual (English, German,

French and Japanese): Order 070-9898-02, Service Manual: Order 070-9899-01,

Carrying Case: Order 016-0156-06.

2