Characteristics, Recommended oscilloscopes, Datasheet characteristics – Atec Tektronix-TCP0150 User Manual
Page 2
Datasheet
Characteristics
Warranted
Characteristic
Description
Bandwidth
DC to ≥20 MHz (-3 dB). Derate with Frequency. See
Figure 1. (Calculated from Rise Time BW = .35/Rise
Time)
Rise Time
≤17.5 ns
Max RMS Current
150 A
Max Peak Pulse Current
500 A
Minimum Sensitivity
5 mA (on oscilloscope’s 1 mV/div setting)
AC Coupling (on oscilloscopes that support AC/DC
coupling)
Current Ranges
25 A and 150 A
Max Bare-wire Voltage
600 V
RMS
CAT I and II
300 V
RMS
CAT III
Typical
Characteristic
Description
DC Accuracy
±1% Typical
3% Warranted
Max Amp-second Product
25 A Range
3,000 Aµs
150 A Range
15,000 Aµs
0.03 Ω at 1 MHz
0.075 Ω at 5 MHz
.125 at 10 MHz
Insertion Impedance
.3 at 20 MHz
Signal Delay
≈ 21 ns
Displayed RMS Noise
≤500 μA
RMS
(probe only)
Environmental
Characteristic
Description
Temperature
Operating
0 ºC to +50 ºC
Nonoperating
-40 ºC to +75 ºC
Humidity
Operating
5% to 95% RH (Relative Humidity) at up to +30 ºC
5% to 85% RH above 30 ºC up to +50 ºC,
noncondensing
Nonoperating
5% to 95% RH (Relative Humidity) at up to +30 ºC
5% to 85% RH above 30 ºC up to +75 ºC,
noncondensing
Altitude
Operating
Up to 2,000 meters (6,560 feet)
Nonoperating
Up to 12,000 meters (40,000 feet)
Physical Characteristics
Dimension
Meters
Inches
Cable Length
2 meters
79 in.
Probe head size
Length
26.8 cm
10.55 in.
Width
4.1 cm
1.6 in.
Height
15.6 cm
6.13 in.
Max Conductor Diameter
21 mm x 25 mm
0.83 in. x 1.0 in.
Weight
kg
lb.
Shipping
.7865
1.734
Power Requirements
TCP0150 is powered directly by oscilloscopes
with the TekVPI interface
Recommended Oscilloscopes
Oscilloscopes with the TekVPI probe interface.
Note: For best probe support, download and install the latest version of the
oscilloscope software from
2
www.tektronix.com