Characteristics, Data sheet – Atec Tektronix-TCPA Series User Manual
Page 2
Data Sheet
Meets Today's AC/DC Current Measurement
Applications
The TCPA300 amplifier, when used with TCP312, TCP305, or TCP303
probes, provides a wide range of current measurement capability and spans
the gap between low-level milliamp measurements to very high current
levels. These three probes provide current measurement capabilities of
30 A, 50 A, and 150 A DC continuous. For even higher current levels, the
TCPA400 amplifier with the TCP404XL current probe measures 500 A DC
continuous and 750 A DC continuous, derated with duty cycle.
Higher frequency performance is available with the TCP312 w/TCPA300
providing ≥100 MHz bandwidth and a maximum current of 30 A DC.
Measurement Errors and Manual Calculations are Now
a Thing of the Past
With this new series of current measurement tools, automatic control and
on-screen scaling and units is provided for users of Tektronix TDS3000,
TDS500, TDS600, TDS700, TDS5000, TDS6000, and TDS7000B Series
oscilloscope systems (the DPO3000, DPO4000, and DPO7000 Series
oscilloscopes, the TPA-BNC adapter is required).
The TCP300/TCP400 current measurement systems seamlessly integrate
with your TDS Series oscilloscope.
Even non-TEKPROBE systems can use the TCPA300/400 Series to make
proper current measurements by simply multiplying the measured output
voltage on the oscilloscope by the TCPA300/400 Series range setting.
Characteristics
Characteristic
TCP312 w/ TCPA300
TCP305 w/ TCPA300
TCP303 w/ TCPA300
TCP404XL w/ TCPA400
Bandwidth (–3 dB)
DC – 100 MHz
DC – 50 MHz
DC – 15 MHz
DC – 2 MHz*
3
Rise Time
≤3.5 ns
≤7 ns
≤23 ns
≤175 ns
Maximum Current Ratings:
High-current Sensitivity Range
10 A/V Range
10 A/V Range
50 A/V Range
1 A/mV Range
DC (Continuous)
30 A
50 A
150 A
500 A (750 A*
4
)
RMS (Sinusoidal)
21.2 A
35.4 A
150 A
500 A
Peak Pulse
50 A
50 A
500 A
750 A
Low-current Sensitivity Range
1 A/V Range
5 A/V Range
5 A/V Range
N/A
DC (Continuous)
5 A
25 A
25 A
N/A
RMS (Sinusoidal)
3.5 A
17.7 A
17.7 A
N/A
Peak Pulse
50 A
50 A
500 A
N/A
DC Accuracy
(Operating temp 0 °C to 50 °C)
±3% of reading
±3% of reading
±3% of reading (10 °C to 50 °C)
+3%/–6% of reading (0 °C
to <10 °C)
±3% of reading
DC Accuracy, Typical
(Operating temp 23 °C ±5 °C)
±1% of reading
±1% of reading
±1% of reading
±1% of reading
Nominal
Maximum Bare Wire Voltage
For use with insulated wires only
600 V
RMS
CAT I & II
300 V
RMS
CAT III
1 mA
5 mA
5 mA
1 A
Lowest Measurable Current
(at ±3% accuracy at DC)
Scope set to 1 mV/div and 20 MHz BW Limited
Scope set to 1 mV/div and 20 MHz BW Limited
Insertion Impedance
(See curves below)
0.08 Ω at 1 MHz
0.15 Ω at 10 MHz
0.27 Ω at 50 MHz
0.7 Ω at 100 MHz
0.035 Ω at 1 MHz
0.12 Ω at 10 MHz
0.4 Ω at 50 MHz
0.01 Ω at 1 MHz
0.025 Ω at 5 MHz
0.1 Ω at 15 MHz
0.1 mΩ at 10 kHz
0.6 mΩ at 100 kHz
8 mΩ at 1 MHz
16 mΩ at 2 MHz
2
www.tektronix.com