Atec Victoreen-660 Series User Manual
Page 2
APPLICATIONS
The 660 probes make use of ion chamber technology, ideal for use in standard
radiology, fluoroscopy, and mammography quality assurance programs where
inbeam
exposure and exposure rate measurements over a broad range of kVp energies
are required. The 660-5 Scatter Probe is designed primarily for Health Physics
exposure assessments of scattered radiation.
FEATURES
• Circular parallel plate ion chambers have assured atmospheric communication
• The 660 family of chambers also maintain precision within 1% short
term and 2% per year at 20°C exclusive of air density or energy response
• The 660 series converts the ion chamber current to a pulse train thereby
eliminating effects from flexing cables
• Ambient Dose Equivalent versions of these probes have energy response
curves per the ICRP recommendations
SPECIFICATIONS
Series Number
660-3
660-3-DE
660-4A
660-4A-DE
660-5
660-5-DE
Application
Diagnostic Beam
Measurement
Diagnostic Beam
Measurement
Scatter
Probe Volume
4cc
4cc
400 cc
Measurement Area 10 cm2
10 cm2
100 cm2
Maximum Rate
999 R/min
99.9 R/min
9.99 R/hr
Maximum Exposure
99.9 R
99.9 R
99.9 mR
Resolution on most
sensitive range
Rate
10 mR/min
1 mR/min
0.1 mR/hr
Resolution on most
sensitive range
Exposure
1 mR
0.1 mR
1 µR
Intensity Limit for
99% Collection
Efficiency
4 R/sec
1.8 R/sec
10.8 R/hr