Atec Victoreen-660 Series User Manual

Page 2

background image

APPLICATIONS
The 660 probes make use of ion chamber technology, ideal for use in standard
radiology, fluoroscopy, and mammography quality assurance programs where
inbeam
exposure and exposure rate measurements over a broad range of kVp energies
are required. The 660-5 Scatter Probe is designed primarily for Health Physics
exposure assessments of scattered radiation.

FEATURES
• Circular parallel plate ion chambers have assured atmospheric communication
• The 660 family of chambers also maintain precision within 1% short
term and 2% per year at 20°C exclusive of air density or energy response
• The 660 series converts the ion chamber current to a pulse train thereby
eliminating effects from flexing cables
• Ambient Dose Equivalent versions of these probes have energy response
curves per the ICRP recommendations

SPECIFICATIONS
Series Number

660-3
660-3-DE

660-4A
660-4A-DE

660-5
660-5-DE

Application

Diagnostic Beam
Measurement

Diagnostic Beam
Measurement

Scatter

Probe Volume

4cc

4cc

400 cc

Measurement Area 10 cm2

10 cm2

100 cm2

Maximum Rate

999 R/min

99.9 R/min

9.99 R/hr

Maximum Exposure

99.9 R

99.9 R

99.9 mR

Resolution on most
sensitive range
Rate

10 mR/min

1 mR/min

0.1 mR/hr

Resolution on most
sensitive range
Exposure

1 mR

0.1 mR

1 µR

Intensity Limit for
99% Collection
Efficiency

4 R/sec

1.8 R/sec

10.8 R/hr