beautypg.com

Technical data - dm70 – Atec Vaisala-DM70 User Manual

Page 2

background image

100 Commerce Way, Woburn, MA 01801

Tel: 1-888-VAISALA (824-7252)

Fax: (781) 933-8029

Email: [email protected]

Online: www.vaisala-usa.com

DMP74 Dewpoint Measurement Probe

Measured variables
Dewpoint

Measurement range (typical)

-76…+140 °F (-60...+60 °C)

Accuracy -50...+60 °C

±3.6 °F (±2 °C)

see the graph below

Response time

63% (90%) at flow rate 0.2 m/s, 1 bar pressure, +68 °F (+20 °C)

-58 to 32 °F T

d

(-50 to 0 °C T

d

)

10 s [35 s]

32 to -4 °F T

d

(0 to -20 °C T

d

)

10 s [35 s]

32 to -40 °F T

d

(0 to -40 °C T

d

)

15 s [120 s]

32 to -58 °F T

d

(0 to -50 °C T

d

)

35 s [300 s]

Temperature

Measurement range

14 °F …140 °F (-10....+60 °C)

Accuracy at +68 °F (+20°C)

±0.36 °F (±0.2 °C)

Typical temperature dependence of electronics

±0.005 °C/°C

Temperature sensor

Pt100 IEC751 1/3 class B

ppm volume concentration

Measurement range (typical)

40...200 000 ppm

Accuracy at +68 °F (+20°C)

7.3 ppm + 8.3% of reading

Relative humidity

Measurement range

0...100 %RH

Accuracy at +68 °F (+20°C)

RH < 10%

±0.025 %RH + 8.75% of reading

RH > 10%

±0.7 %RH + 2% of reading

Measurement environment

Temperature

14 °F …140 °F (-10....+60 °C)

Pressure

0...20 bar

Sample flow rate

no effect

Probe general

Sensor

DRYCAP

‚

180S

Probe material (wetted parts)

Stainless steel (AISI 316L)

Sensor protection

Sintered filter (AISI 316L)

part no: HM47280

Mechanical connection

G1/2" ISO228-1 thread

with bonded seal ring (U-seal)

Housing classification

IP65 (NEMA4)

Weight

350 g

MI70 Measurement Indicator

Indicator general

Menu languages

English, German, French,

Finnish and Spanish

Display

LCD with backlight

Graphic trend display of any parameter

Character height up to 16 mm

Probe inputs

1 or 2

Power Supply

Rechargeable NIMH battery pack

with AC-POWER or

4xA4 size alkalines, type IEC LR6

Analog output

0...1 VDC

Output resolution

0.6 mV

Data interface

RS232C

Data-logging capacity

900...2700 points with real time

Alarm

Audible alarm function

Operating temperature range

14 °F …104 °F (-10....+40 °C)

Operating humidity range

non-condensing

Housing classification

IP54

Weight

400 g

Battery operation time

Continuous use

48 h typical at +68 °F (+20°C)

Data logging use

up to 30 days

DMP74 Probe + MI70 Indicator = DM70 Hand-held

Dewpoint Meter

General

Housing materials

ABS/PC blend

Storage temperature

-40 …158 °F (-40...+70 °C)

Storage humidity

0..100 %RH non condensing

Total weight

750 g

Electromagnetic compatibility

Complies with the following standard:

EN 61326-1:1997 + Am1:1998, Electrical equipment for

measurement, control and laboratory use - EMC requirements

- Generic Environment.

Accessories and DM70KIT

DSC74 sample cell adapter

sampling system for pressurized processes

Pressure range

0...10 bar

g

Weight

300 g

Connection cables

for DMT242 Transmitter

order no: 27160ZZ

for DMP248 Transmitter

order no: 27159ZZ

Analog output cable

order no: 27168ZZ

Windows

®

software

including PC connection cable

order no: MI70LINK

Carrying Case

MI70Case

DM70 Calibration Kit

order no: DM70KIT

includes DM70 indicator with rechargeable battery and 115 VAC

adapter; MIK70 link & analog output signal cable; DMP74 probe;

DSC74 Sample cell and quick adapter; carrying case; certificate of

calibration, traceable to NIST.

Specifications subject to change without prior notice. DRYCAP® is

a registered trademark of Vaisala. ©Vaisala Oyj

TECHNICAL DATA - DM70

De

wpoint temper

ature ( C)

0

20

40

60

0

40

-60

-80

Process temperature ( C)

10

20

30

50

60

70

-10
-20
-30

-50

-40

-70

30

10

50

Dewpoint meas

urement range

lower limit

-10

Accurac

y < ± 2

C Td