Technical data - dm70 – Atec Vaisala-DM70 User Manual
Page 2
![background image](/manuals/319861/2/background.png)
100 Commerce Way, Woburn, MA 01801
Tel: 1-888-VAISALA (824-7252)
Fax: (781) 933-8029
Email: [email protected]
Online: www.vaisala-usa.com
DMP74 Dewpoint Measurement Probe
Measured variables
Dewpoint
Measurement range (typical)
-76 +140 °F (-60...+60 °C)
Accuracy -50...+60 °C
±3.6 °F (±2 °C)
see the graph below
Response time
63% (90%) at flow rate 0.2 m/s, 1 bar pressure, +68 °F (+20 °C)
-58 to 32 °F T
d
(-50 to 0 °C T
d
)
10 s [35 s]
32 to -4 °F T
d
(0 to -20 °C T
d
)
10 s [35 s]
32 to -40 °F T
d
(0 to -40 °C T
d
)
15 s [120 s]
32 to -58 °F T
d
(0 to -50 °C T
d
)
35 s [300 s]
Temperature
Measurement range
14 °F 140 °F (-10....+60 °C)
Accuracy at +68 °F (+20°C)
±0.36 °F (±0.2 °C)
Typical temperature dependence of electronics
±0.005 °C/°C
Temperature sensor
Pt100 IEC751 1/3 class B
ppm volume concentration
Measurement range (typical)
40...200 000 ppm
Accuracy at +68 °F (+20°C)
7.3 ppm + 8.3% of reading
Relative humidity
Measurement range
0...100 %RH
Accuracy at +68 °F (+20°C)
RH < 10%
±0.025 %RH + 8.75% of reading
RH > 10%
±0.7 %RH + 2% of reading
Measurement environment
Temperature
14 °F 140 °F (-10....+60 °C)
Pressure
0...20 bar
Sample flow rate
no effect
Probe general
Sensor
DRYCAP
180S
Probe material (wetted parts)
Stainless steel (AISI 316L)
Sensor protection
Sintered filter (AISI 316L)
part no: HM47280
Mechanical connection
G1/2" ISO228-1 thread
with bonded seal ring (U-seal)
Housing classification
IP65 (NEMA4)
Weight
350 g
MI70 Measurement Indicator
Indicator general
Menu languages
English, German, French,
Finnish and Spanish
Display
LCD with backlight
Graphic trend display of any parameter
Character height up to 16 mm
Probe inputs
1 or 2
Power Supply
Rechargeable NIMH battery pack
with AC-POWER or
4xA4 size alkalines, type IEC LR6
Analog output
0...1 VDC
Output resolution
0.6 mV
Data interface
RS232C
Data-logging capacity
900...2700 points with real time
Alarm
Audible alarm function
Operating temperature range
14 °F 104 °F (-10....+40 °C)
Operating humidity range
non-condensing
Housing classification
IP54
Weight
400 g
Battery operation time
Continuous use
48 h typical at +68 °F (+20°C)
Data logging use
up to 30 days
DMP74 Probe + MI70 Indicator = DM70 Hand-held
Dewpoint Meter
General
Housing materials
ABS/PC blend
Storage temperature
-40 158 °F (-40...+70 °C)
Storage humidity
0..100 %RH non condensing
Total weight
750 g
Electromagnetic compatibility
Complies with the following standard:
EN 61326-1:1997 + Am1:1998, Electrical equipment for
measurement, control and laboratory use - EMC requirements
- Generic Environment.
Accessories and DM70KIT
DSC74 sample cell adapter
sampling system for pressurized processes
Pressure range
0...10 bar
g
Weight
300 g
Connection cables
for DMT242 Transmitter
order no: 27160ZZ
for DMP248 Transmitter
order no: 27159ZZ
Analog output cable
order no: 27168ZZ
Windows
®
software
including PC connection cable
order no: MI70LINK
Carrying Case
MI70Case
DM70 Calibration Kit
order no: DM70KIT
includes DM70 indicator with rechargeable battery and 115 VAC
adapter; MIK70 link & analog output signal cable; DMP74 probe;
DSC74 Sample cell and quick adapter; carrying case; certificate of
calibration, traceable to NIST.
Specifications subject to change without prior notice. DRYCAP® is
a registered trademark of Vaisala. ©Vaisala Oyj
TECHNICAL DATA - DM70
De
wpoint temper
ature ( C)
0
20
40
60
0
40
-60
-80
Process temperature ( C)
10
20
30
50
60
70
-10
-20
-30
-50
-40
-70
30
10
50
Dewpoint meas
urement range
lower limit
-10
Accurac
y < ± 2
C Td