beautypg.com

Options – Atec Tenney-T2 Series User Manual

Page 2

background image

Model

T2RC

T2C

Workspace
inches-(mm)

W

17 (432)

17 (432)

D

13 (330)

13 (330)

H

16 (406)

16 (406)

Exterior
inches-(mm)

W

26 (660)

26 (660)

D

26 (660)

26 (660)

H

67 (1702)

65 (1651)

Temperature Range & Humidity

Low

°C

-75

-75

High

°C

+200

+200

Change Rates in Minutes

Ambient to °C
Chamber Empty

200°C

90 Minutes

90

150°C

40 Minutes

40

100°C

25 Minutes

25

93°C

17 Minutes

17

65°C

9 Minutes

9

2°C

5 Minutes

5

-7°C

7 Minutes

7

-12°C

10 Minutes

10

-40°C

30 Minutes

30

-54°C

45 Minutes

45

-65°C

70 Minutes

70

-73°C

100 Minutes

100

-75°C

Ultimate

Ultimate

Utilities, etc

Refrigeration

(2)1/2 HP + 1/2 HP

(2)1/2 HP + 1/2 HP

Heater Capacity

500 W

500 W

Humidifier

W

500

-

GPH

0.15

-

AMPS @ 115V, 1Ph

18

18

AMPS Fuse

20

20

Unit Weight

LBS/Kg

350/158

350/158

Options:

• IEEE/488 interface
• LinkTenn software for Windows

®

that permits your

computer to control up to 10 chambers

• RS-422, 423, 232 or 485 interface assemblies
• Upgrade to VersaTenn III (standard on T2RC)
• Water reservoir for humidity system (5 gallon)
• Recirculating system for humidity water
• Additional ports
• Event relay board
• Water demineralizer
• Viewing window, 6” x 7”, thermally insulated and

heated (no manual wiper required)

• Interior light
• Shelving, adjustable and removable
• Automatic CO2 or LN2 cooling boost system
• Automating boost heating system
• GN2 purge system
• Recording instruments
• Redundant thermal protection and alarm system
• External dryer for obtaining humidity as low as 5%

(to 20°C)

• Alternate power supply wiring

* Humidity capablility: 20% to 98% RH in the dry bulb range of +20°C to 85°C as limited by a 3°C dewpoint. Test data based on

24°C ambient, sea level, 60Hz.

* On 50 Hz or higher than 24°C ambient, performance may be reduced. Consult factory regarding any special cooling requirements.

CFC-free refrigerants are used exclusively on all Tenney chambers.

Gruenberg, Blue M, Tenney, Lunaire
Mailing Address: P.O. Box 150 | White Deer, PA 17887-0150 | USA |
Phone: (570) 538-7200 | Fax: (570) 538-7380
Physical Address: 2821 Old Route 15 | New Columbia, PA 17856-9396
[email protected]

tps-tenney-t2-series-temperature-and-humidity-cycling-chambers-20100129