beautypg.com

Tenney-tjr-tujr-thjr_datasheet.pdf, Specifications – Atec Tenney-TJR-TUJR-THJR User Manual

Page 2

background image

Specifications

Model

TJR

TUJR

THJR

Workspace

W

D
H

Exterior

W

D
H

Overall dimensions in inches/centimeters

Low

°C

High

°C

Temperature Range

200°

55°

150°

30

100°

15

93°

30

65°

15

20

-7°

30

-12°

45

-40°

20

-54°

30

-65°

40

-73°

55

-75°

Ultimate

Change rates in minutes

Ambient to °C
to Chamber
Empty

15

e

/40

11/28

11

:

/30

36

:

/93

28/71

51/130

22

2

/57

20/51

24/61

30

2

/78

65/165

30/76

16

2

/42

15

:

/40

20/51

-75

+200

-12

+93

24°

275

10°

170

-7°

45

-12°

5

-40°

170

-54°

145

-65°

105

-73°

60

-75°

Ultimate

Live load capacity in watts (humidity system off)

Temperature °C

Refrigeration

(2)

2

HP +

a

HP

(1)

1

5

HP

Heater Capacity

500 W

500 W

Humidifier

W

500

GPH

0.15

AMPS @ 115V, 1

Ø

18

12

AMPS Fuse

20

15

Unit Weight

LBS/Kg

260/118

275/125

—

—

—

—

Utilities, etc.

Humidity capability: 20% to 98% RH in the dry bulb range of +20ºC to +85ºC as limited by a 3º dewpoint. Test data based on 24ºC ambient, sea level, 60Hz. On 50 Hz

or higher than 24ºC ambient, performance may be reduced. Consult factory regarding any special cooling requirements. CFC-free refrigerants are used exclusively on

all Tenney chambers.