beautypg.com

Se-1000 series specifications cascade chambers, Full range humidity specifications, 1000 series humidity range – Atec Thermotron-SE-1000 Series User Manual

Page 2

background image

SE-1000 SERIES SPECIFICATIONS

CASCADE CHAMBERS

Interior Dimensions — W x D x H

Inches / Centimeters

All Models 40 x 39.25 x 38.25 / 102 x 100 x 97

Volume — Cubic Ft / Liters

All Models 34.8 / 986

Exterior Dimensions — W x D x H

Inches / Centimeters

All Models 49 x 83 x 87 / 125 x 211 x 221

Temperature Range

All Models 180ºC to -70ºC (356ºF to -94ºF)

Temperature Control

All Models ±0.3ºC (±0.5ºF)

Temperature Uniformity

1

All Models ±0.5ºC (±0.9ºF)

Airflow

All Models 1,000 cfm

Window Size — W x H

Inches / Centimeters

All Models 15 x 19 / 38 x 48

Compressor Size — Hp

Two 3.5

Two 6

Two 10

Two 15

Noise Level

2

— dBA

Heating / Cooling

60 / 74

60 / 76

60 / 76

60 / 76

Approximate
Shipping Weight — Lbs / Kg

1,840 / 834

1,900 / 862

2,340 / 1,064 2,415 / 1,098

Cooling Performance Capacity — Minutes

Air Temperature (Empty Chamber)
180ºC to -65ºC (356ºF to -85ºF)

78

56

34

27

71ºC to -65ºC (160ºF to -85ºF)

50

36

21

16

85ºC to -40ºC (185ºF to -40ºF)

35

26

13

11

Cooling Product Temperature — Minutes

50Lb / 23Kg Aluminum
71ºC to -65ºC (160ºF to -85ºF)

68

53

32

28

85ºC to -40ºC (185ºF to -40ºF)

53

46

26

24

Heating Performance Capacity — Minutes

Air Temperature (Empty Chamber)
-65ºC to 180ºC (-85ºF to 356ºF)

43

43

21

12

-65ºC to 71ºC (-85ºF to 160ºF)

20

20

10

-40ºC to 85ºC (-40ºF to 185ºF)

19

19

9

7

Heating Product Temperature — Minutes

50Lb / 23Kg Aluminum
-65ºC to 71ºC (-85ºF to 160ºF)

44

44

27

25

-40ºC to 85ºC (-40ºF to 185ºF)

41

41

26

24

Live Load Capacity — Watts, Temp. Mode

0ºC (32ºF)

2500

3000

3000

3000

-54ºC (-65.2ºF)

1500

2000

2000

2000

Electrical Service — No Humidity / Humidity

208 / 3 / 60 (Full Load Amps)

54 / 58

71 / 71

92 / 92

150 / —

230 / 3 / 60 (Full Load Amps)

51 / 57

65 / 65

83 / 83

136 / —

460 / 3 / 60 (Full Load Amps)

26 / 29

32 / 32

42 / 42

68 / —

400 / 3 / 50 (Running Load Amps)

24 / 27

33 / 33

45 / 45

70 / —

Inlet Water — Gal / Liters per Minute

29ºC / 85ºF

Air-Cooled

Air-Cooled

16 / 60

24 / 91

24ºC / 75ºF

Air-Cooled

Air-Cooled

9 / 34

14 / 53

18ºC / 65ºF

Air-Cooled

Air-Cooled

7.5 / 28

11 / 42

13ºC / 55ºF

Air-Cooled

Air-Cooled

6 / 23

8 / 31

Model

SE-1000-3-3

SE-1000-6-6 SE-1000-10-10 SE-1000-15-15

SE-1000 SERIES SPECIFICATIONS

SINGLE STAGE CHAMBERS

Interior Dimensions — W x D x H

Inches / Centimeters

All Models 40 x 39.25 x 38.25 / 102 x 100 x 97

Volume — Cubic Ft / Liters

All Models 34.8 / 986

Exterior Dimensions — W x D x H

Inches / Centimeters

All Models 49 x 83 x 87 / 125 x 211 x 221

Temperature Range

All Models 180ºC to -40ºC (356ºF to -40ºF)

Temperature Control

All Models ±0.3ºC (±0.5ºF)

Temperature Uniformity

1

All Models ±0.5ºC (±0.9ºF)

Airflow

All Models 1,000 cfm

Window Size — W x H

Inches / Centimeters

All Models 15 x 19 / 38 x 48

Compressor Size — Hp

4

6

10

15

Noise Level

2

— dBA

Heating / Cooling

60 / 74

60 / 76

60 / 76

60 / 76

Approximate
Shipping Weight — Lbs / Kg

1,660 / 753

1,860 / 844

2,065 / 937

2,115 / 959

Cooling Performance Capacity — Minutes

Air Temperature (Empty Chamber)
180ºC to -35ºC (356ºF to -31ºF)

74

52

37

34

71ºC to -35ºC (160ºF to -31ºF)

46

33

22

21

85ºC to -20ºC (185ºF to -4ºF)

32

22

12

10

Cooling Product Temperature — Minutes

50Lb / 23Kg Aluminum
71ºC to -35ºC (160ºF to -31ºF)

66

49

36

35

85ºC to -20ºC (185ºF to -4ºF)

48

39

29

28

Heating Performance Capacity — Minutes

Air Temperature (Empty Chamber)
-35ºC to 180ºC (-31ºF to 356ºF)

36

36

20

10

-35ºC to 71ºC (-31ºF to 160ºF)

15

15

8

5

-20ºC to 85ºC (-4ºF to 185ºF)

15

15

8

5

Heating Product Temperature — Minutes

50Lb / 23Kg Aluminum
-35ºC to 71ºC (-31ºF to 160ºF)

33

33

25

23

-20ºC to 85ºC (-4ºF to 185ºF)

33

33

25

23

Live Load Capacity — Watts, Temp. Mode

0ºC (32ºF)

2500

3000

3000

3000

-20ºC (-4ºF)

1500

2000

2000

2000

Electrical Service — No Humidity / Humidity

208 / 3 / 60 (Full Load Amps)

39 / 55

45 / 61

60 / 72

95 / 106

230 / 3 / 60 (Full Load Amps)

37 / 55

42 / 60

55 / 68

86 / 99

460 / 3 / 60 (Full Load Amps)

19 / 28

21 / 30

27 / 33

43 / 50

400 / 3 / 50 (Running Load Amps)

19 / 28

24 / 32

40 / 40

50 / 55

Inlet Water — Gal / Liters per Minute

29ºC / 85ºF

Air-Cooled

Air-Cooled

16 / 60

24 / 91

24ºC / 75ºF

Air-Cooled

Air-Cooled

9 / 34

14 / 53

18ºC / 65ºF

Air-Cooled

Air-Cooled

7.5 / 28

11 / 42

13ºC / 55ºF

Air-Cooled

Air-Cooled

6 / 23

8 / 31

Model

SE-1000-4

SE-1000-5

SE-1000-10

SE-1000-15

THERMOTRON INDUSTRIES

291 Kollen Park Drive
Holland, Michigan, USA 49423
Mktg: (616) 393-4580
Main: (616) 392-1491
Fax: (616) 392-5643
E-mail: [email protected]

THERMOTRON INDUSTRIES, U.K.

Newton House
Winch Road
Kent Science Park
Sittingbourne, Kent
ME9 8EF England
Phone: 01795 436333
Fax: 01795 436777
Email: [email protected]

Visit us on the Internet

www.thermotron.com

FORM NO. T04-005 09/04 (REV. 04/06) ©THERMOTRON INDUSTRIES PRINTED IN USA
UNITED STATES PATENT NO. 6,976,340 and 5,631,429

A full range, modular humidity system can be added to SE-Series Test Chambers
either at the time of purchase or as a field retrofit. Performance specifications for
the humidity module are characterized by a wide range, precise uniformity and
tight control.

1 Standard deviation from mean, measured at -25°C (-13°F) or at +100°C (+212°F).

2 Noise Level: A-weighted sound pressure level measured at a distance of 1.0 meter

(39.4 inches) from the surface of the equipment at a height of 1.6 meters
(63 inches) from the floor in free-field conditions, using a calibrated instrument.

3 Limited by a +7°C (+45°F) minimum dew point temperature and

a maximum dry bulb temperature of +88°C (+190°F).

4 At a dry bulb temperature above +20°C (+68°F).

5 Based upon temperature uniformity specifications.

Relative humidity indication at or near the physical limits may be affected by sensor
accuracy and control tolerance. An optional humidity package can be added for
applications requiring humidity levels lower than those covered by the full range
humidity system.

FULL RANGE HUMIDITY SPECIFICATIONS*

Humidity Range

3

10% to 98% RH

Dry Bulb Temperature Range

7°C to 88°C (45°F to 190°F)

Dewpoint Temperature Range

7°C to 87°C (45°F to 188°F)

Extended Dewpoint Condition

-10°C (14°F)

Humidity Control

4

±2.5% RH

Humidity Uniformity

5

±1.0% RH

* Except 15hp Cascade Chamber models

STANDARD HUMIDITY RANGE

-10°C (14°F) DEWPOINT

Performance is based upon 60 Hz and laboratory ambient conditions of 23.9°C (75°F), and may vary slightly on 50 Hz power or at other
temperature and humidity levels. Chambers are designed for use under normal laboratory operating conditions. For other applications, please
consult Thermotron.

It is Thermotron’s understanding that the unit will be used in a non-hazardous environment. The unit is not designed for use with or for the
purpose of processing hazardous materials. If hazardous materials are involved, please consult the factory for an alternative quote for a properly
designed unit.

The addition of accessories may impact performance.

Specifications subject to change without notice.

1000 SERIES HUMIDITY RANGE

100

80

60

40

20

0

0

20

40

60

80

100°C

32

68

104

140

176

212°F

PERCENT

RH