S-3800 series temperature chamber specifications, Sm-3800 series humidity range – Atec Thermotron-S-3800_S-SM User Manual
Page 4
THERMOTRON INDUSTRIES
291 Kollen Park Drive
Holland, Michigan, USA 49423
Mktg: (616) 393-4580
Main: (616) 392-1491
Fax: (616) 392-5643
THERMOTRON INDUSTRIES, U.K.
Newton House
Winch Road
Sittingbourne Research Centre
Sittingbourne, Kent
ME9 8EF England
Phone: 01795 436333
Fax: 01795 436777
Email: [email protected]
Visit us on the Internet
www.thermotron.com
E-mail: info
@
thermotron.com
FORM NO. T04-001 04/04 ©THERMOTRON INDUSTRIES
PRINTED IN USA
S-3800 SERIES TEMPERATURE CHAMBER SPECIFICATIONS
Model
S-4 3800
S-8 3800
S-16 3800
S-27 3800
S-32 3800
Temperature Range
All Models -70º to +180ºC (-94º to +356ºF)
Cooling Change Rate — Min
+180º to -65ºC (+356º to -85ºF)
50
75
75
95
70
+71º to -65ºC (+160º to -85ºF)
35
52
55
70
48
+85º to -40ºC (+185º to -40ºF)
24
36
38
50
30
Heating Change Rate — Min
-65º to +180ºC (-85º to +356ºF)
36
42
54
73
46
-65º to +71ºC (-85º to +160ºF)
15
17
22
31
20
-40º to +85ºC (-40º to +185ºF)
13
16
21
30
18
Temperature Control Tolerance**
All Models
±0.3ºC (±0.5ºF)
Live Load Capacity — Watts
-18ºC (0ºF)
600
550
1000
1000
1000
-40ºC (-40ºF)
400
350
700
700
700
-54ºC (-65ºF)
300
200
400
400
400
Temperature Uniformity**
All Models
±0.7ºC ( ±1.1ºF)
Window — In/Cm
12x12/30x30
13x19/33x48
17x23/43x58
12x12/30x30
17x23/43x58
Cubic Volume — Ft/Liters
4/113
8/227
16/453
27/764
32/906
Weight — Lbs/Kg
700/318
800/363
1320/599
1800/817
1800/817
Workspace — WxDxH
Inches
20x20x20
24x24x24
30x30x30
36x36x36
38x38x38
Centimeters
51x51x51
61x61x61
76x76x76
91x91x91
97x97x97
Move-in — WxDxH
Inches
31x42x68
35x49x73
45x59x81
51x65x88
53x68x90
Centimeters
79x107x173
89x124x185
114x150x206
130x166x224
135x173x229
Electrical Service — Amps
(full load)
(full load)
(full load)
(full load)
(full load)
230/1/60
25
25
47
47
47
230/3/60
20
20
36
36
36
220/1/50
—
28
—
—
—
400/3/50
17
17
18
18
18
Noise Level† — dBA
Heating/Cooling†
60/68
60/68
60/70
60/70
60/70
SM-3800 SERIES TEMPERATURE / HUMIDITY CHAMBER SPECIFICATIONS
Model
SM-8 3800
SM-16 3800
SM-27 3800
SM-32 3800
Temperature Range
All Models -68º to +180ºC (-90º to +356ºF)
Humidity Range*
All Models 10% to 98% RH
Cooling Change Rate — Min
+180º to -65ºC (+356º to -85ºF)
78
85
102
74
+71º to -65ºC (+160º to -85ºF)
54
59
76
51
+85º to -40ºC (+185º to -40ºF)
38
41
53
31
Heating Change Rate — Min
-65º to +180ºC (-85º to +356ºF)
43
56
76
48
-65º to +71ºC (-85º to +160ºF)
17
22
33
20
-40º to +85ºC (-40º to +185ºF)
16
21
31
18
Temperature Control Tolerance**
All Models
±0.3ºC (±0.5ºF)
Live Load Capacity — Watts
-18ºC (0ºF)
550
1000
1000
1000
-40ºC (-40ºF)
350
700
700
700
-54ºC (-65ºF)
200
400
400
400
Temperature Uniformity**
All Models
±0.7ºC ( ±1.1ºF)
Humidity Control Tolerance***
All Models 2.5% RH
Humidity Uniformity****
All Models
±1% RH
Window — In/Cm
13x19/33x48
17x23/43x58
12x12/30x30
17x23/43x58
Cubic Volume — Ft/Liters
8/227
16/453
27/764
32/906
Weight — Lbs/Kg
875/397
1395/633
1875/851
1875/851
Workspace — WxDxH
Inches
24x24x24
30x30x30
36x36x36
38x38x38
Centimeters
61x61x61
76x76x76
91x91x91
97x97x97
Move-in — WxDxH
Inches
35x49x73
45x59x81
51x65x88
53x68x90
Centimeters
89x124x185
114x150x206
130x166x224
135x173x229
Electrical Service — Amps
(full load)
(full load)
(full load)
(full load)
230/1/60
25
50
50
50
230/3/60
20
37
37
37
220/1/50
28
—
—
—
400/3/50
17
29
29
29
Noise Level† — dBA
Heating/Cooling†
60/68
60/70
60/70
60/70
* Limited by a +7
°C (+45°F) minimum dew point
temperature and a maximum dry bulb
temperature of +88
°C (+190°F). Relative
humidity indication at or near the physical
limits may be affected by sensor accuracy
and control tolerance.
** Standard deviation from mean, measured
at -25
°C (-13°F) or at +100°C (+212°F).
*** At a dry bulb temperature above +20
°C (+68°F).
**** Based upon temperature uniformity
specifications.
† Noise Level: A-weighted sound pressure level
measured at a distance of 1.0 meter (39.4 inches)
from the surface of the equipment at a height of
1.6 meters (63 inches) from the floor in free-field
conditions, using a calibrated instrument.
Performance is based upon 60 Hz and 23.9
°C (75°F)
ambient air, and may vary slightly at other ambient
temperatures. Chambers are designed for use under
normal laboratory operating conditions. For other
applications, please consult Thermotron.
SM-3800 SERIES HUMIDITY RANGE
ULTRA LOW TEMPERATURE / HUMIDITY CHAMBER
0
20
40
60
80
100
°C
32
68
104
140
176
212
°F
0
20
40
60
80
100
°C
32
68
104
140
176
212
°F
100
80
60
40
20
0
WITH OPTIONAL LOW HUMIDITY PACKAGE
100
80
60
40
20
0
PERCENT RH
PERCENT RH