S-3800 SERIES TEMPERATURE CHAMBER SPECIFICATIONS

Model

S-4 3800

S-8 3800

S-16 3800

S-27 3800

S-32 3800

Temperature Range

All Models -70º to +180ºC (-94º to +356ºF)

Cooling Change Rate — Min

+180º to -65ºC (+356º to -85ºF)

50

75

75

95

70

+71º to -65ºC (+160º to -85ºF)

35

52

55

70

48

+85º to -40ºC (+185º to -40ºF)

24

36

38

50

30

Heating Change Rate — Min

-65º to +180ºC (-85º to +356ºF)

36

42

54

73

46

-65º to +71ºC (-85º to +160ºF)

15

17

22

31

20

-40º to +85ºC (-40º to +185ºF)

13

16

21

30

18

Temperature Control Tolerance**

All Models

±0.3ºC (±0.5ºF)

Live Load Capacity — Watts

-18ºC (0ºF)

600

550

1000

1000

1000

-40ºC (-40ºF)

400

350

700

700

700

-54ºC (-65ºF)

300

200

400

400

400

Temperature Uniformity**

All Models

±0.7ºC ( ±1.1ºF)

Window — In/Cm

12x12/30x30

13x19/33x48

17x23/43x58

12x12/30x30

17x23/43x58

Cubic Volume — Ft/Liters

4/113

8/227

16/453

27/764

32/906

Weight — Lbs/Kg

700/318

800/363

1320/599

1800/817

1800/817

Workspace — WxDxH

Inches

20x20x20

24x24x24

30x30x30

36x36x36

38x38x38

Centimeters

51x51x51

61x61x61

76x76x76

91x91x91

97x97x97

Move-in — WxDxH

Inches

31x42x68

35x49x73

45x59x81

51x65x88

53x68x90

Centimeters

79x107x173

89x124x185

114x150x206

130x166x224

135x173x229

Electrical Service — Amps

(full load)

(full load)

(full load)

(full load)

(full load)

230/1/60

25

25

47

47

47

230/3/60

20

20

36

36

36

220/1/50

28

400/3/50

17

17

18

18

18

Noise Level† — dBA

Heating/Cooling†

60/68

60/68

60/70

60/70

60/70

SM-3800 SERIES TEMPERATURE / HUMIDITY CHAMBER SPECIFICATIONS

Model

SM-8 3800

SM-16 3800

SM-27 3800

SM-32 3800

Temperature Range

All Models -68º to +180ºC (-90º to +356ºF)

Humidity Range*

All Models 10% to 98% RH

Cooling Change Rate — Min

+180º to -65ºC (+356º to -85ºF)

78

85

102

74

+71º to -65ºC (+160º to -85ºF)

54

59

76

51

+85º to -40ºC (+185º to -40ºF)

38

41

53

31

Heating Change Rate — Min

-65º to +180ºC (-85º to +356ºF)

43

56

76

48

-65º to +71ºC (-85º to +160ºF)

17

22

33

20

-40º to +85ºC (-40º to +185ºF)

16

21

31

18

Temperature Control Tolerance**

All Models

±0.3ºC (±0.5ºF)

Live Load Capacity — Watts

-18ºC (0ºF)

550

1000

1000

1000

-40ºC (-40ºF)

350

700

700

700

-54ºC (-65ºF)

200

400

400

400

Temperature Uniformity**

All Models

±0.7ºC ( ±1.1ºF)

Humidity Control Tolerance***

All Models 2.5% RH

Humidity Uniformity****

All Models

±1% RH

Window — In/Cm

13x19/33x48

17x23/43x58

12x12/30x30

17x23/43x58

Cubic Volume — Ft/Liters

8/227

16/453

27/764

32/906

Weight — Lbs/Kg

875/397

1395/633

1875/851

1875/851

Workspace — WxDxH

Inches

24x24x24

30x30x30

36x36x36

38x38x38

Centimeters

61x61x61

76x76x76

91x91x91

97x97x97

Move-in — WxDxH

Inches

35x49x73

45x59x81

51x65x88

53x68x90

Centimeters

89x124x185

114x150x206

130x166x224

135x173x229

Electrical Service — Amps

(full load)

(full load)

(full load)

(full load)

230/1/60

25

50

50

50

230/3/60

20

37

37

37

220/1/50

28

400/3/50

17

29

29

29

Noise Level† — dBA

Heating/Cooling†

60/68

60/70

60/70

60/70

* Limited by a +7

°C (+45°F) minimum dew point

temperature and a maximum dry bulb
temperature of +88

°C (+190°F). Relative

humidity indication at or near the physical
limits may be affected by sensor accuracy
and control tolerance.

** Standard deviation from mean, measured

at -25

°C (-13°F) or at +100°C (+212°F).

*** At a dry bulb temperature above +20

°C (+68°F).

**** Based upon temperature uniformity

specifications.

† Noise Level: A-weighted sound pressure level

measured at a distance of 1.0 meter (39.4 inches)
from the surface of the equipment at a height of
1.6 meters (63 inches) from the floor in free-field
conditions, using a calibrated instrument.

Performance is based upon 60 Hz and 23.9

°C (75°F)

ambient air, and may vary slightly at other ambient
temperatures. Chambers are designed for use under
normal laboratory operating conditions. For other
applications, please consult Thermotron.

SM-3800 SERIES HUMIDITY RANGE

ULTRA LOW TEMPERATURE / HUMIDITY CHAMBER

0

20

40

60

80

100

°C

32

68

104

140

176

212

°F

0

20

40

60

80

100

°C

32

68

104

140

176

212

°F

100

80

60

40

20

0

WITH OPTIONAL LOW HUMIDITY PACKAGE

100

80

60

40

20

0

PERCENT RH

PERCENT RH