Options – Atec Tenney-BTC-BTRC User Manual
Page 2
Humidity capability: 20% to 98% RH in the dry bulb range of +20°C (68°F) to +85°C (185°F)
as limited by a 3 degree dew point.
Options:
• IEEE/488 interface
• LinkTenn software for Windows that permits
your computer to control up to 10 chambers
• RS-422, 423, 232, or 485 interface assemblies
• Water demineralizer
• Water reservoir for humidity system (5 gallon)
• Recirculating system for humidity water
• Viewing window, 6” x 8”, thermally insulated
and heated
• Interior light
• Shelving, adjustable and removable
• Automatic CO2 or LN2 cooling boost system
• Automating boost heating system
• GN2 purge system
• Recording instruments
• Redundant thermal protection and alarm system
• External dryer for obtaining humidity as low as
5% (to 20°C)
• Alternate power supply wiring
• Chamber cart with casters
Model
BTC
BTRC
Workspace
W
20/51
D
19.25/49
H
22/56
Exterior
W
57/145
62 /157
D
34/86
H
34.5/88
Temperature Range
Low
°C
-73
High
°C
200
Utilities, etc.
Refrigeration
(2) 1 HP
Heater Capacity
2 KW
Humidifi er
Watts
750
GPH
0.3
AMPS@230V, 1 Ph
24
24
AMPS, Fuse
35
Unit Weight
LBS
500
Overall dimensions in inches/centimeters
Gruenberg, Blue M, Tenney, Lunaire
Mailing Address: P.O. Box 150 | White Deer, PA 17887-0150 | USA |
Phone: (570) 538-7200 | Fax: (570) 538-7380
Physical Address: 2821 Old Route 15 | New Columbia, PA 17856-9396
[email protected]
tps-tenney-benchmaster-temperature-and-humidity-environmental-chamber-20100129