NESLAB ThermoFlex
™
900 Specifications
Setpoint Temperature Range
5°C - 40°C
41°F - 104°F
Ambient Temperature Range
10°C - 40°C
50°F - 104°F
Temperature Stability
+/- 0.1°C
Setpoint Cooling Capacity at 20°C (see graph below for cooling curves)
60 Hz
900W (3072 BTU)
50 Hz
750W (2560 BTU)
Reservoir Volume
Gallons 1.9
Liters 7.6
Footprint or dimensions (HxWxD)
Inches
27.4 x 14.2 x 24.7
cm
69.6 x 36.1 x 62.7
Unit Weight
lb 126
kg 57.2
Pumps (see graph below for performance curves)
PD 1 - Positive Displacement
60 Hz
2.1 gpm @ 60 psi
50 Hz
1.7 gpm @ 60 psi
PD 2 - Positive Displacement
60 Hz
4.1 gpm @ 60 psi
50 Hz
3.3 gpm @ 60 psi
Power options
Option 1
100V/50Hz/1 Phase 15A Circuit
Option 2
100V/60Hz/1 Phase 15A Circuit
Option 3
115V/60Hz/1 Phase 15A Circuit
Option 4
208-230V/60Hz/1 Phase 15A Circuit
Option 5
230V/50Hz/1 Phase Amp 16A Circuit
Compliance
NRTL Certified to CSA and
UL Standards; CE-Marked
Product Specfications
Cooling Capacity
Pumping Capacity
Specifications obtained at sea level using water as the recirculating fluid, at a 20°C process setpoint, 25°C ambient condition, at nominal
operating voltage. Other fluids, fluid temperatures, ambient temperatures, altitude or operating voltages will affect performance. Thermo Electron
reserves the right to change specifications without notice. Pump performance results obtained with no restrictions on return to the system.