A d v a n c e d Te s t S o l u t i o n s f o r E M C
Parameter Value
Voltage Range
0.2 to 4.4 kV
Oscillation frequency
100 kHz and 1 MHz
Pulse repetition
1 MHz pulse: From 1/s to 600/s
1)
, default is 400/s
100 kHz pulse: From 1/s to 120/s
1)
, default is 40/s
Source impedance
150 Ω and 200 Ω
Burst duration
From 1 s to 100 s or 1 pulse to 9999 pulses or continuous –
default setting is 2 s
Rise time
75 ns
EUT AC
1 Ph 16 A, 24 to 270 Vrms, 50/60 Hz (phase - neutral), 400 Hz max.
EUT DC
16 A, 0 to 270 VDC
Coupling cap
0.5 µF
Decoupling choke
2 x ≤1.5 mH
Output Floating
Options
Type Description
WIN 3000
Control and reporting software
MD 200
Voltage calibration probe
MD 300-SOW
Current calibration probe
CDN 3425
Capacitive coupling clamp
INA XXX
Adapter for capacitive coupling clamp
INA 166
Rack mounting brackets (4HU)
© June 2013 Teseq
®
Specifications subject to change without notice.
Teseq
®
is an ISO-registered company. Its products
are designed and manufactured under the strict
quality and environmental requirements of the ISO
9001. This document has been carefully checked.
However, Teseq
®
does not assume any liability for
errors or inaccuracies.
Teseq AG
Nordstrasse 11F 4542 Luterbach Switzerland
T + 41 32 681 40 40 F + 41 32 681 40 48
[email protected]
www.teseq.com
691-309A June 2013
THE SMART 4 KV SOLUTION FOR
SLOW DAMPED OSCILLATORY WAVE TESTING
NSG 3040-SOW
Slow Damped Oscillatory Wave pulses 100 kHz and 1 MHz
Pulse conforms to IEC/EN 61000-4-18, ANSI C37.90.1, and IEC/EN 62052-11
The NSG 3040-SOW performs tests according to the following specifi cations:
1)
Pulse repetition rate is derated for voltage levels above 3 kV. Please refer to user manual.
Dimensions
449 (17.7”) x 226 (8.9”; 5 HU) x 565 mm (22.2”) (W x H x D)
Weight
approx. 25 kg (55 lbs)
Dimensions / weight