Model e513, Specifications and tolerances, Available options – Atec Thermo Scientific-E513 User Manual
Page 2
SPECIFICATIONS AND TOLERANCES
W
AVEFORMS
Voltage Ramps:
0.1kV/µs, 0.5kV/µs, 1.0kV/µs,
5.0kV/µs, 10kV/µs, 0.1kV/µs is
linear to 2.5kV; all other ramps
linear to 3.0kV
Note: Specified ramp rates are
obtained with an open-circuit
voltage setting of 3.0kV.
Voltage
~65µs for 0.1kV/µs;
Durations:
~40µs for 0.5kV/µs and 1kV/µs;
~5µs for 5kV/µs and 10kV/µs
Current
~45µs at 0.1kV/µs;
Durations:
~40µs at 0.5kV/µs and 1.0kV/µs;
~5µs at 5kV/µs and 10kV/µs
Open-Circuit
0-3000V
Voltage:
±5% in 1 volt steps
Short-Circuit
50A, ±10% when the
Current:
peak open-circuit voltage
is set to 3.0kV
NOTE:
To obtain linear fronts, waves are
quasi-square waves with 20-25%
initial overshoots beyond peak
open-circuit voltages, except for the
0.1kV/µs which is roughly triangular.
Undershoots range from 5 to 25%.
AVAILABLE OPTIONS
E513-VI:
Provides monitoring of the peak surge
voltages and currents at the output of
the E513 module. All measurements
are logged by software for diagnostic
evaluation or Go/No-Go testing.
Note: If an ECAT coupler/decoupler
is included, waveform monitoring is
available at the output of the
coupler/decoupler without the
addition of Option E513-VI.
E513-S:
Adds an oscilloscope trigger for any
Surge Network.
Model E513
ECAT: AN EXTRAORDINARY CONCEPT IN EMC
TESTING FOR VULNERABILITY TO SURGE,
EFT AND POWER QUALITY.
ECAT is a totally integrated and modularized system
which offers unprecedented testing ease, accuracy and
versatility to manufacturers, test houses, and anyone
who will be testing for effects of Pulsed EMI.
ECAT is designed to test products according to a broad
range of international standards developed by ANSI,
the IEC, CENELEC and other agencies. The European
Norms, developed by CENELEC, became mandatory
on January 1, 1996 for all equipment shipped into the
European Union or shipped across EU borders.
KPS-520B-11/97
A division of Thermo Voltek, a Thermo Electron company
KeyTek and ECAT are registered trademarks and SurgeWare and PQF are trademarks
of Thermo Voltek Corporation. ©Thermo Voltek. Specifications are subject to change
without notice. Printed in USA.
One Lowell Research Center
Lowell, Massachusetts 01852-4345 USA
1 800 753 9835 • Tel: 1 978 275 0800 • Fax: 1 978 275 0850
email: [email protected]
http://www.keytek.com
For additional ECAT
®
product literature call, email or
fax the KeyTek sales department.
P
HYSICAL
Height:
21cm (8
1
/
4
”)
Width:
21cm (8
1
/
4
”)
Depth:
52cm (20
3
/
8
”)
Weight:
53kg (24 lbs.)
M
INIMUM
S
YSTEM
R
EQUIREMENTS
E100 series control center